Calibration of a two-color soft x-ray diagnostic for electron temperature measurement.
Reusch, L M; Den Hartog, D J; Franz, P; Goetz, J; McGarry, M B; Stephens, H D.
Afiliação
  • Reusch LM; University of Wisconsin - Madison, Madison, Wisconsin 53703, USA.
  • Den Hartog DJ; University of Wisconsin - Madison, Madison, Wisconsin 53703, USA.
  • Franz P; Consorzio RFX, Padova, Italy.
  • Goetz J; University of Wisconsin - Madison, Madison, Wisconsin 53703, USA.
  • McGarry MB; University of Wisconsin - Madison, Madison, Wisconsin 53703, USA.
  • Stephens HD; University of Wisconsin - Madison, Madison, Wisconsin 53703, USA.
Rev Sci Instrum ; 87(11): 11E332, 2016 Nov.
Article em En | MEDLINE | ID: mdl-27910453
The two-color soft x-ray (SXR) tomography diagnostic on the Madison Symmetric Torus is capable of making electron temperature measurements via the double-filter technique; however, there has been a 15% systematic discrepancy between the SXR double-filter (SXRDF) temperature and Thomson scattering (TS) temperature. Here we discuss calibration of the Be filters used in the SXRDF measurement using empirical measurements of the transmission function versus energy at the BESSY II electron storage ring, electron microprobe analysis of filter contaminants, and measurement of the effective density. The calibration does not account for the TS and SXRDF discrepancy, and evidence from experiments indicates that this discrepancy is due to physics missing from the SXRDF analysis rather than instrumentation effects.
Buscar no Google
Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies / Prognostic_studies Idioma: En Ano de publicação: 2016 Tipo de documento: Article
Buscar no Google
Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies / Prognostic_studies Idioma: En Ano de publicação: 2016 Tipo de documento: Article