Reliability of Pulse Oximetry during Progressive Hypoxia, Cardiopulmonary Resuscitation, and Recovery in a Piglet Model of Neonatal Hypoxic Cardiac Arrest.
Hassan, Mohammad Ahmad; Weber, Claudia; Waitz, Markus; Huang, Li; Hummler, Helmut D; Mendler, Marc Robin.
Afiliação
  • Hassan MA; Department of Pediatrics, Faculty of Medicine, Sohag University, Sohag, Egypt.
Neonatology ; 112(1): 40-46, 2017.
Article em En | MEDLINE | ID: mdl-28253519
ABSTRACT

BACKGROUND:

Pulse oximetry is widely used in intensive care and emergency conditions to monitor arterial oxygenation and to guide oxygen therapy.

OBJECTIVE:

To study the reliability of pulse oximetry in comparison with CO-oximetry in newborn piglets during progressive hypoxia, cardiac arrest, cardiopulmonary resuscitation (CPR), and after return of spontaneous circulation (ROSC).

METHODS:

Thirty-three newborn piglets were exposed to hypoxia until asystole occurred and then resuscitated until ROSC. Arterial oxygen saturation was monitored continuously by pulse oximetry (SpO2) with one sensor applied to the wrist of the right forelimb (FL) and another to the thigh of the left hind limb (HL). Arterial functional oxygen saturation (SaO2) was measured at baseline and at predefined intervals during each phase of the experiment. SpO2 was compared with coinciding SaO2 values and bias considered whenever the difference (SpO2 - SaO2) was beyond ±5%.

RESULTS:

Bias values were lower at the baseline measurements (-3.7 ± 2.3% in FL and -4.1 ± 3.4% in HL) as well as after ROSC (1.5 ± 4.2% in FL and 0.2 ± 4.6% in HL) with higher precision and accuracy than during other experiment phases. During hypoxia induction, cardiac arrest, and CPR, there was a marked decrease in precision and accuracy as well as an increase in bias up to 43 ± 26 and 56 ± 27% in FL and HL, respectively, over a range of SaO2 from 13 to 51%.

CONCLUSION:

Pulse oximetry showed increased bias and decreased accuracy and precision during marked hypoxemia in a model of neonatal hypoxic cardiac arrest.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Oxigênio / Oximetria / Reanimação Cardiopulmonar / Membro Anterior / Parada Cardíaca / Membro Posterior / Hipóxia Tipo de estudo: Etiology_studies / Prognostic_studies Limite: Animals Idioma: En Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Oxigênio / Oximetria / Reanimação Cardiopulmonar / Membro Anterior / Parada Cardíaca / Membro Posterior / Hipóxia Tipo de estudo: Etiology_studies / Prognostic_studies Limite: Animals Idioma: En Ano de publicação: 2017 Tipo de documento: Article