Your browser doesn't support javascript.
loading
Subcutaneous cardioverter defibrillator has longer time to therapy but is less cardiotoxic than transvenous cardioverter defibrillator. Study carried out in a preclinical porcine model.
Garcia, Rodrigue; Inal, Sofiane; Favreau, Frederic; Jayle, Christophe; Hauet, Thierry; Bruneval, Patrick; Kerforne, Thomas; Hajj-Chahine, Jamil; Degand, Bruno.
Afiliação
  • Garcia R; CHU Poitiers, Service de Cardiologie, 2 rue de la Milétrie, Poitiers 86021, France.
  • Inal S; Univ Poitiers, Faculté de Médecine et de Pharmacie, Poitiers 86000, France.
  • Favreau F; Inserm U1082, Poitiers, 86000, France.
  • Jayle C; CHU de Poitiers, Service de Biochimie, Poitiers 86000, France.
  • Hauet T; Univ Poitiers, Faculté de Médecine et de Pharmacie, Poitiers 86000, France.
  • Bruneval P; Inserm U1082, Poitiers, 86000, France.
  • Kerforne T; CHU de Poitiers, Service de Biochimie, Poitiers 86000, France.
  • Hajj-Chahine J; Univ Poitiers, Faculté de Médecine et de Pharmacie, Poitiers 86000, France.
  • Degand B; Inserm U1082, Poitiers, 86000, France.
Europace ; 20(5): 873-879, 2018 05 01.
Article em En | MEDLINE | ID: mdl-28460030
ABSTRACT

Aims:

Totally subcutaneous implantable cardioverter defibrillator (S-ICD) delivers higher shock energy and can have longer time to therapy compared to transvenous implantable cardioverter defibrillator (T-ICD). Aim of the study was to compare time to therapy and to investigate cardiac, cerebral and systemic injuries of S-ICD and T-ICD shocks delivered after ventricular fibrillation (VF) induction. Methods and

results:

Fourteen pigs were randomly implanted with a S-ICD (n = 7) or a T-ICD (n = 7). Five VF episodes were induced in each pig. For each VF episode, up to two shocks could be delivered by the T-ICD or the S-ICD to terminate the arrhythmia. Cardiac, systemic, and cerebral toxicity were monitored. Mean time to therapy was longer in the S-ICD group compared to the T-ICD group (19[18; 23] s vs. 9 [7; 10] s; P = 0.001, respectively). High-sensitivity troponin T levels were significantly higher in the T-ICD group from 1 to 24 h after the procedure (P ≤ 0.02). Creatine phosphokinase activity levels were significantly higher in the S-ICD group, at 3, 6, and 24 h after the procedure (P ≤ 0.05). Lactate levels were not significantly different between groups. S100 protein level was similar in both groups at 1 h after the procedure and then decreased in the T-ICD group compared to the S-ICD group (P = 0.04).

Conclusions:

Time to therapy in S-ICD was twice as long as for T-ICD, but didn't induce relevant brain injury. Conversely, S-ICD shocks were less cardiotoxic than T-ICD shocks.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Fibrilação Ventricular / Cardioversão Elétrica / Desfibriladores Implantáveis Limite: Animals Idioma: En Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Fibrilação Ventricular / Cardioversão Elétrica / Desfibriladores Implantáveis Limite: Animals Idioma: En Ano de publicação: 2018 Tipo de documento: Article