Evaluation of the effect of temperature variation on response of PRESAGE® dosimeter.
Farhood, Bagher; Khezerloo, Davood; Zadeh, Tohid Morteza; Nedaie, Hassan Ali; Hamrahi, Daryoosh; Khezerloo, Najimeh.
Afiliação
  • Farhood B; Department of Medical Physics and Biomedical Engineering, School of Medicine, Tehran University of Medical Sciences, Tehran, Iran.
  • Khezerloo D; Department of Medical Physics and Biomedical Engineering, School of Medicine, Tehran University of Medical Sciences, Tehran, Iran.
  • Zadeh TM; Department of Medical Physics and Biomedical Engineering, School of Medicine, Tehran University of Medical Sciences, Tehran, Iran.
  • Nedaie HA; Department of Medical Physics and Biomedical Engineering, School of Medicine, Tehran University of Medical Sciences, Tehran, Iran.
  • Hamrahi D; Sadra Radiotherapy Center, Qom, Iran.
  • Khezerloo N; Department of Radiology, Imam Hospital, Tabriz University of Medical Sciences, Tabriz, Iran.
J Cancer Res Ther ; 13(1): 118-121, 2017.
Article em En | MEDLINE | ID: mdl-28508844
ABSTRACT

INTRODUCTION:

Many factors, such as PRESAGE ® composition, dose rate, energy, and type of radiation, temperature, etc., may effect on PRESAGE ® dosimeter response. The aim of this study was investigating the effect of temperature variation on response of PRESAGE ® solid dosimeter. MATERIALS AND

METHODS:

In this study, a PRESAGE ® solid detector was fabricated. Ninety-four percent weight polyurethane, 5% weight carbon tetrachloride, and 1% weight leucomalachite green were used. Radiological and physical characteristics of PRESAGEs ®, such as mass density, electron density, and effective number atomic were obtained and compared with water. Response of PRESAGE ® dosimeter in temperatures -4, 10, 25, 35, 45, 55, 65, 75, 85, and 90°C was evaluated. In addition, the absorption peak at various temperatures was investigated.

RESULTS:

The results showed that the absorption peak at different temperatures was in the range of 630-635 nm. For temperatures below 75°C, the results indicated that temperature variation has no effect on the response of PRESAGE ® dosimeter whereas at the temperatures> 75°C, temperature variation has an effect on PRESAGE ® dosimeter response.

CONCLUSION:

The finding showed that temperature changes have not impact on the absorption peak. In addition, the results related to the effect of temperature variation on the response of PRESAGE ® dosimeter showed that in the range of clinical applications (temperatures below 75°C), temperature variation has no effect on PRESAGE ® dosimeter response.