3D electrochemical and ion current imaging using scanning electrochemical-scanning ion conductance microscopy.
Takahashi, Yasufumi; Ida, Hiroki; Matsumae, Yoshiharu; Komaki, Hirokazu; Zhou, Yuanshu; Kumatani, Akichika; Kanzaki, Makoto; Shiku, Hitoshi; Matsue, Tomokazu.
Afiliação
  • Takahashi Y; WPI-Advanced Institute for Materials Research, Tohoku University, 980-8577, Japan. matsue@bioinfo.che.tohoku.ac.jp.
Phys Chem Chem Phys ; 19(39): 26728-26733, 2017 Oct 11.
Article em En | MEDLINE | ID: mdl-28951914
Local cell-membrane permeability and ionic strength are important factors for maintaining the functions of cells. Here, we measured the spatial electrochemical and ion concentration profile near the sample surface with nanoscale resolution using scanning electrochemical microscopy (SECM) combined with scanning ion-conductance microscopy (SICM). The ion current feedback system is an effective way to control probe-sample distance without contact and monitor the kinetic effect of mediator regeneration and the chemical concentration profile. For demonstrating 3D electrochemical and ion concentration mapping, we evaluated the reaction rate of electrochemical mediator regeneration on an unbiased conductor and visualized inhomogeneous permeability and the ion concentration 3D profile on a single fixed adipocyte cell surface.

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2017 Tipo de documento: Article