Ocular anatomic changes for different accommodative demands using swept-source optical coherence tomography: a pilot study.
Esteve-Taboada, José J; Ferrer-Blasco, Teresa; Aloy, Miguel A; Adsuara, José E; Cerdá-Durán, Pablo; Mimica, Petar; Montés-Micó, Robert.
Afiliação
  • Esteve-Taboada JJ; Department of Optics and Optometry and Visual Science, University of Valencia, C/ Dr Moliner, 50, Burjassot, 46100, Spain. josejuan.esteve@uv.es.
  • Ferrer-Blasco T; Department of Optics and Optometry and Visual Science, University of Valencia, C/ Dr Moliner, 50, Burjassot, 46100, Spain.
  • Aloy MA; Department of Astronomy and Astrophysics, University of Valencia, C/ Dr Moliner, 50, Burjassot, 46100, Spain.
  • Adsuara JE; Department of Astronomy and Astrophysics, University of Valencia, C/ Dr Moliner, 50, Burjassot, 46100, Spain.
  • Cerdá-Durán P; Department of Astronomy and Astrophysics, University of Valencia, C/ Dr Moliner, 50, Burjassot, 46100, Spain.
  • Mimica P; Department of Astronomy and Astrophysics, University of Valencia, C/ Dr Moliner, 50, Burjassot, 46100, Spain.
  • Montés-Micó R; Department of Optics and Optometry and Visual Science, University of Valencia, C/ Dr Moliner, 50, Burjassot, 46100, Spain.
Graefes Arch Clin Exp Ophthalmol ; 255(12): 2399-2406, 2017 Dec.
Article em En | MEDLINE | ID: mdl-28951958
ABSTRACT

PURPOSE:

The purpose of our study was to assess the changes in ocular parameters for different accommodative demands using a new optical biometer based on swept-source optical coherence tomography (SS-OCT).

METHODS:

Seven subjects were included in this pilot study, and only one eye per participant was analyzed. Each eye was measured six times with the optical biometer IOLMaster 700 (Carl Zeiss Meditec, Jena, Germany). As this instrument is not able to change the vergence of the stimulus, to enable measurements at different accommodative states, a tilted first-surface mirror attached to the optical biometer was used to place the fixation stimulus at different vergences. Measurements were taken on the right eye of the subject while the left eye was looking through the mirror. Central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), axial length (AL), white-to-white (WTW) distance and keratometric readings were evaluated for three different accommodative states 0.0 diopters (D), 1.5 D, and 3.0 D.

RESULTS:

No statistically significant differences were found for CCT, AL, WTW, K1 and K2 between the three accommodative states. As expected, changing the accommodative condition did not change CCT, AL, WTW, and keratometric outcomes. Nevertheless, statistically significant differences between the accommodative states were found for ACD and LT measurements. In addition, variations in ACD correlated linearly with variations in LT (R2 ≥ 0.99) when changing the vergence of the optotype.

CONCLUSION:

A practical methodology to assess the changes in ocular parameters for different accommodative demands using the IOLMaster 700 based on SS-OCT has been described. Statistically significant changes that have been found that agree well with previous reports.