Performance of a segmented HPGe detector at KRISS.
Appl Radiat Isot
; 134: 177-181, 2018 Apr.
Article
em En
| MEDLINE
| ID: mdl-28982543
ABSTRACT
A 24 segmented HPGe coaxial detector was set up with a digitized data acquisition system (DAQ). The DAQ was composed of a digitizer (5 × 107sampling/s), a Field-Programmable Gate Array (FPGA), and a real time operating system. The Full Width Half Maximum (FWHM), rise time, signal characteristics, and spectra of a 137Cs source were evaluated. The data were processed using an in-house developed gamma-ray tracking system.