Bioprosthesis are commonly used in the elderly population. Structural valve deterioration affects the long-term durability. We report an early deterioration of a Mitroflow valve caused by ring deformation and prosthetic leaflet rupture. The 69-years-old patient underwent successful redo surgery with excision of the bioprosthesis and placement of a mechanical valve.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Estenose Aórtica Subvalvar / Valva Aórtica / Bioprótese / Falha de Prótese / Próteses Valvulares Cardíacas / Implante de Prótese de Valva Cardíaca Limite: Aged / Humans / Male Idioma: En Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Estenose Aórtica Subvalvar / Valva Aórtica / Bioprótese / Falha de Prótese / Próteses Valvulares Cardíacas / Implante de Prótese de Valva Cardíaca Limite: Aged / Humans / Male Idioma: En Ano de publicação: 2018 Tipo de documento: Article