Sensing performance, safety, and patient acceptability of long-dipole cardiac monitor: An innovative axillary insertion.
Pacing Clin Electrophysiol
; 41(3): 277-283, 2018 03.
Article
em En
| MEDLINE
| ID: mdl-29341174
BACKGROUND: The recommended location for implantable cardiac monitor (ICM) insertion is the left pectoral region. We tested whether an innovative left axillary implantation approach could be applicable for a new ICM, characterized by a long sensing dipole. METHODS: We considered a series of 55 patients consecutively implanted with a long-dipole ICM (BioMonitor 2); the first 30 subjects underwent prepectoral location insertion, while the subsequent 25 received the ICM in the axillary region. Sensing performances collected at 1-month follow-up were compared between the two groups. During the visit, each patient was also asked to fill in a brief questionnaire to assess patient acceptability of the device. RESULTS: All patients had a successful insertion of ICM. Mean R-wave amplitude was 0.87 ± 0.44 mV in the prepectoral group and 1.00 ± 0.45 mV in the axillary one, without any significant difference. The percentage of patients with visible P wave was also comparable between the two approaches (65.5% vs 68.2%, P = 0.84). None of the patients reported device-related issues or discomfort, and ICM was generally well accepted and tolerated by all the involved patients. CONCLUSION: Axillary insertion may represent a valid alternative to the standard one for long-dipole ICM technology providing not only patient acceptability but also high-quality sensing performances.
Palavras-chave
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Axila
/
Aceitação pelo Paciente de Cuidados de Saúde
/
Eletrocardiografia Ambulatorial
/
Implantação de Prótese
/
Segurança do Paciente
Limite:
Aged
/
Female
/
Humans
/
Male
/
Middle aged
Idioma:
En
Ano de publicação:
2018
Tipo de documento:
Article