Development of stereotactic radiosurgery using carbon beams (carbon-knife).
Phys Med Biol
; 63(4): 045024, 2018 02 20.
Article
em En
| MEDLINE
| ID: mdl-29364137
The aim of this research is to develop a stereotactic-radiosurgery (SRS) technique using carbon beams to treat small intracranial lesions; we call this device the carbon knife. A 2D-scanning method is adapted to broaden a pencil beam to an appropriate size for an irradiation field. A Mitsubishi slow extraction using third order resonance through a rf acceleration system stabilized by a feed-forward scanning beam using steering magnets with a 290 MeV/u initial beam energy was used for this purpose. Ridge filters for spread-out Bragg peaks (SOBPs) with widths of 5 mm, 7.5 mm, and 10 mm were designed to include fluence-attenuation effects. The collimator, which defines field shape, was used to reduce the lateral penumbra. The lateral-penumbra width at the SOBP region was less than 2 mm for the carbon knife. The penumbras behaved almost the same when changing the air gap, but on the other hand, increasing the range-shifter thickness mostly broadened the lateral penumbra. The physical-dose rates were approximate 6 Gy s-1 and 4.5 Gy s-1 for the 10 × 10 mm2 and 5 × 5 mm2 collimators, respectively.