Radiotherapy and risk of implantable cardioverter-defibrillator malfunctions: experimental data from direct exposure at increasing doses.
Zecchin, Massimo; Artico, Jessica; Morea, Gaetano; Severgnini, Mara; Bianco, Elisabetta; De Luca, Antonio; Fantasia, Anna Zorzin; Salvatore, Luca; Milan, Vittorino; Lucarelli, Matteo; Dissegna, Roberta; Cannatà, Antonio; Sinagra, Gianfranco.
Afiliação
  • Zecchin M; Dipartimento Cardiovascolare.
  • Artico J; Dipartimento Cardiovascolare.
  • Morea G; Dipartimento Cardiovascolare.
  • Severgnini M; Dipartimento di Fisica Sanitaria.
  • Bianco E; Dipartimento Cardiovascolare.
  • De Luca A; Dipartimento Cardiovascolare.
  • Fantasia AZ; Dipartimento Cardiovascolare.
  • Salvatore L; Dipartimento Cardiovascolare.
  • Milan V; Dipartimento di Radioterapia, Azienda Sanitaria Universitaria Integrata di Trieste, Trieste, Italy.
  • Lucarelli M; Dipartimento Cardiovascolare.
  • Dissegna R; Dipartimento Cardiovascolare.
  • Cannatà A; Dipartimento Cardiovascolare.
  • Sinagra G; Dipartimento Cardiovascolare.
J Cardiovasc Med (Hagerstown) ; 19(4): 155-160, 2018 04.
Article em En | MEDLINE | ID: mdl-29389805
ABSTRACT

AIM:

During radiotherapy, in patients with implantable cardioverter-defibrillators (ICDs) malfunctions are considered more likely if doses more than 2 Gy reach the ICD site; however, most malfunctions occur with high-energy (>10 MV) radiations, and the risk is less defined using 6-MV linear accelerators. The purpose of the study is to experimentally evaluate the occurrence of malfunctions in ICDs radiated with a 6-MV linear accelerator at increasing photon doses.

METHODS:

Thirty-two ICDs from all manufacturers (31 explanted and one demo) were evaluated; all devices with a sufficient battery charge underwent multiple radiations with a 6-MV photon beam reaching a cumulative dose at ICD site of 0.5, 1, 2, 3, 5 and 10 Gy and interrogated after every session. All antitachycardia therapies were left enabled; two ICDs were connected to a rhythm simulator (one simulating a complete atrioventricular block without ventricular activity) and visually monitored by external ECG and the ICD programmer during radiation.

RESULTS:

Thirteen ICDs were excluded before radiation because of battery depletion; after radiation up to the cumulative dose at the cardiac implantable electronic device site of 10 Gy, in the remaining 19 devices, programmation and battery charge remained unchanged and no switch to safety mode was observed; oversensing, pacing inhibition or inappropriate antitachycardia therapy were neither recorded nor visually observed during radiation.

CONCLUSION:

With a low-energy accelerator, neither malfunctions nor electromagnetic interferences were detected radiating the ICDs at doses usually reaching the ICD pocket during radiotherapy sessions. In this context, magnet application to avoid oversensing and inappropriate therapy seems, therefore, useless.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Radioterapia / Desfibriladores Implantáveis / Análise de Falha de Equipamento Tipo de estudo: Etiology_studies / Risk_factors_studies Limite: Humans Idioma: En Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Radioterapia / Desfibriladores Implantáveis / Análise de Falha de Equipamento Tipo de estudo: Etiology_studies / Risk_factors_studies Limite: Humans Idioma: En Ano de publicação: 2018 Tipo de documento: Article