Evaluation of the UFXC32k photon-counting detector for pump-probe experiments using synchrotron radiation.
J Synchrotron Radiat
; 25(Pt 2): 413-418, 2018 Mar 01.
Article
em En
| MEDLINE
| ID: mdl-29488920
ABSTRACT
This paper presents the performance of a single-photon-counting hybrid pixel X-ray detector with synchrotron radiation. The camera was evaluated with respect to time-resolved experiments, namely pump-probe-probe experiments held at SOLEIL. The UFXC camera shows very good energy resolution of around 1.5â
keV and allows the minimum threshold setting to be as low as 3â
keV keeping the high-count-rate capabilities. Measurements of a synchrotron characteristic filling mode prove the proper separation of an isolated bunch of photons and the usability of the detector in time-resolved experiments.