Dosimetric changes with computed tomography automatic tube-current modulation techniques.
Radiol Phys Technol
; 11(2): 184-191, 2018 Jun.
Article
em En
| MEDLINE
| ID: mdl-29626289
The study is aimed at a verification of dose changes for a computed tomography automatic tube-current modulation (ATCM) technique. For this purpose, anthropomorphic phantom and Gafchromic® XR-QA2 films were used. Radiochromic films were cut according to the shape of two thorax regions. The ATCM algorithm is based on noise index (NI) and three exam protocols with different NI were chosen, of which one was a reference. Results were compared with dose values displayed by the console and with Poisson statistics. The information obtained with radiochromic films has been normalized with respect to the NI reference value to compare dose percentage variations. Results showed that, on average, the information reported by the CT console and calculated values coincide with measurements. The study allowed verification of the dose information reported by the CT console for an ATCM technique. Although this evaluation represents an estimate, the method can be a starting point for further studies.
Palavras-chave
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Tomografia Computadorizada por Raios X
Limite:
Humans
Idioma:
En
Ano de publicação:
2018
Tipo de documento:
Article