Dosing of the second-generation cryoballoon using acute time-to-pulmonary vein isolation as an indicator of durable ablation in a canine model.
Su, Wilber; Coulombe, Nicolas; Kirchhof, Nicole; Grassl, Erin; Wittenberger, Dan.
Afiliação
  • Su W; Banner University Medical Center- Phoenix and University of Arizona, 755 E. McDowell Rd, 4th Floor, Phoenix, AZ, 85006, USA. wilwsu@gmail.com.
  • Coulombe N; Medtronic CryoCath, Pointe-Claire, Canada.
  • Kirchhof N; Medtronic Physiology Research Laboratories, Minneapolis, MN, USA.
  • Grassl E; Medtronic Physiology Research Laboratories, Minneapolis, MN, USA.
  • Wittenberger D; Medtronic CryoCath, Pointe-Claire, Canada.
J Interv Card Electrophysiol ; 53(3): 293-300, 2018 Dec.
Article em En | MEDLINE | ID: mdl-29656325
ABSTRACT

BACKGROUND:

Rigid time-based dosing protocol(s) currently used in the clinic for cryoballoon ablation of atrial fibrillation may be inadequate to guide the circumferential and transmural cryothermal energy transfer across the pulmonary vein (PV) and may result in injury to collateral tissues or electrical gaps between the PV and left atrium (LA).

OBJECTIVE:

A physiologic endpoint (e.g., acute time-to-PV isolation a.k.a. time-to-effect; TTE) may be effective in the determination of a transmural lesion formation and may allow for individualized ablation dosing across each PV.

METHODS:

Thirty PVs from 15 dogs were randomized into five dosing protocols, including (1) TTE + 60 s, (2) TTE + 90 s, (3) TTE + 120 s, (4) TTE + 150 s, and (5) 2 × 180 s. Ablations were conducted with a 23-mm second-generation cryoballoon, and TTE was assessed during a freeze by pacing from an inner balloon-lumen circular diagnostic catheter to a quadripolar diagnostic catheter in the coronary sinus. After ablation, animals were survived for 30 to 34 days, and repeat electrophysiology assessment of PV isolation was conducted after which animals were euthanized for gross anatomy and histological examination.

RESULTS:

At study termination, efficacy endpoint evaluations were based on maintenance of PV electrical isolation, gross anatomy assessment of PV lesions, and histological examination of PVs. Five efficacy endpoint failures were noted, including the following 1 PV in the TTE + 90 sec group; 2 PVs in the TTE + 120 sec group; 1 PV in the TTE + 150 s group; and 1 PV in the 2 × 180 s group. Regarding safety, one phrenic nerve injury was observed in the 2 × 180 s cohort. No other complications were observed.

CONCLUSIONS:

In a canine model, effective PV isolation could be found even in the shortest duration dosing cohort (TTE + 60 s). One complication (phrenic nerve injury) was observed in the longest duration dosing group (2 × 180 s). Further studies will be required to correlate these results to a 28-mm cryoballoon (more commonly used in the cryoablation of a human LA); however, to date, this is the first reporting of a successful cryoablation using TTE + 60 s dosing (approximately 90 s total duration of freezing).
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Veias Pulmonares / Fibrilação Atrial / Criocirurgia / Técnicas Eletrofisiológicas Cardíacas / Átrios do Coração / Complicações Intraoperatórias Tipo de estudo: Etiology_studies / Guideline Limite: Animals Idioma: En Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Veias Pulmonares / Fibrilação Atrial / Criocirurgia / Técnicas Eletrofisiológicas Cardíacas / Átrios do Coração / Complicações Intraoperatórias Tipo de estudo: Etiology_studies / Guideline Limite: Animals Idioma: En Ano de publicação: 2018 Tipo de documento: Article