Your browser doesn't support javascript.
loading
Analysis of SmartPlug Insertion-Related Complications.
Huang, Chien-Jung; Lu, Chia-Ju; Tu, Wen-Hui; Hou, Yu-Chih; Wang, I-Jong; Hu, Fung-Rong; Chen, Wei-Li.
Afiliação
  • Huang CJ; Department of Ophthalmology (C.-J.H., C.-J.L., W.-H.T., Y.-C.H., I.-J.W., F.-R.H., W.-L.C.), National Taiwan University Hospital, Taipei, Taiwan; and Center of Corneal Tissue Engineering and Stem Cell Biology (F.-R.H., W.-L.C.), National Taiwan University Hospital.
Eye Contact Lens ; 44 Suppl 2: S333-S337, 2018 Nov.
Article em En | MEDLINE | ID: mdl-29944494
ABSTRACT

PURPOSE:

To evaluate the complications and the treatment results of SmartPlug-related complications.

METHODS:

Retrospective review of all patients from a single medical hospital who received SmartPlug (Medennium, Inc., Irvine, CA) insertions from October 2007 to February 2014. All patients who developed SmartPlug-related canaliculitis and pyogenic granuloma were analyzed. Lacrimal irrigation with antibiotics was performed in most patients.

RESULTS:

Six hundred thirty-one eyes received SmartPlug insertion and 18 eyes developed SmartPlug-related complications (14 canaliculitis and 4 pyogenic granuloma). The mean time interval from insertion to development of the complications is 3.0 (0.5-6.9) years. Sixteen eyes received lacrimal irrigation with antibiotics, and all the eyes showed improvement without recurrence. The other two eyes had recurrence of granuloma pyogenica only after surgical excision.

CONCLUSIONS:

SmartPlug-related complications, including canaliculitis and granuloma pyogenica, required long-term follow-up. Most of the complications can be cured by lacrimal irrigation of antibiotics.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Complicações Pós-Operatórias / Próteses e Implantes / Síndromes do Olho Seco / Implantação de Prótese Tipo de estudo: Etiology_studies / Observational_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Complicações Pós-Operatórias / Próteses e Implantes / Síndromes do Olho Seco / Implantação de Prótese Tipo de estudo: Etiology_studies / Observational_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2018 Tipo de documento: Article