A cloud-based computer-aided detection system improves identification of lung nodules on computed tomography scans of patients with extra-thoracic malignancies.
Vassallo, Lorenzo; Traverso, Alberto; Agnello, Michelangelo; Bracco, Christian; Campanella, Delia; Chiara, Gabriele; Fantacci, Maria Evelina; Lopez Torres, Ernesto; Manca, Antonio; Saletta, Marco; Giannini, Valentina; Mazzetti, Simone; Stasi, Michele; Cerello, Piergiorgio; Regge, Daniele.
Afiliação
  • Vassallo L; Department of Radiology at Candiolo Cancer Institute-FPO, IRCCS, Strada Provinciale 142 km 3.95, 10060, Candiolo, Turin, Italy. lorenzo.vassallo@ircc.it.
  • Traverso A; Department of Surgical Sciences, University of Turin, A.O.U. Città della Salute e della Scienza, Via Genova 3, 10126, Turin, Italy. lorenzo.vassallo@ircc.it.
  • Agnello M; Department of Applied Science and Technology, Polytechnic University of Turin, Corso Duca Degli Abruzzi 24, 10129, Turin, Italy.
  • Bracco C; Department of Radiation Oncology (MAASTRO), GROW School for Oncology and Developmental Biology, Maastricht University Medical Centre, Dr Tanslaan 12, 6229 ET, Maastricht, The Netherlands.
  • Campanella D; Department of Applied Science and Technology, Polytechnic University of Turin, Corso Duca Degli Abruzzi 24, 10129, Turin, Italy.
  • Chiara G; Medical Physics Unit, Candiolo Cancer Institute-FPO, IRCCS, Strada Provinciale 142 km 3.95, 10060, Candiolo, Turin, Italy.
  • Fantacci ME; Department of Radiology at Candiolo Cancer Institute-FPO, IRCCS, Strada Provinciale 142 km 3.95, 10060, Candiolo, Turin, Italy.
  • Lopez Torres E; Department of Radiology at Candiolo Cancer Institute-FPO, IRCCS, Strada Provinciale 142 km 3.95, 10060, Candiolo, Turin, Italy.
  • Manca A; Pisa Section of INFN, Largo Bruno Pontecorvo 3, 56127, Pisa, Italy.
  • Saletta M; Turin Section of INFN, Via Pietro Giuria 1, 10125, Turin, Italy.
  • Giannini V; Department of Radiology at Candiolo Cancer Institute-FPO, IRCCS, Strada Provinciale 142 km 3.95, 10060, Candiolo, Turin, Italy.
  • Mazzetti S; Turin Section of INFN, Via Pietro Giuria 1, 10125, Turin, Italy.
  • Stasi M; Department of Radiology at Candiolo Cancer Institute-FPO, IRCCS, Strada Provinciale 142 km 3.95, 10060, Candiolo, Turin, Italy.
  • Cerello P; Department of Surgical Sciences, University of Turin, A.O.U. Città della Salute e della Scienza, Via Genova 3, 10126, Turin, Italy.
  • Regge D; Department of Radiology at Candiolo Cancer Institute-FPO, IRCCS, Strada Provinciale 142 km 3.95, 10060, Candiolo, Turin, Italy.
Eur Radiol ; 29(1): 144-152, 2019 Jan.
Article em En | MEDLINE | ID: mdl-29948089
ABSTRACT

OBJECTIVES:

To compare unassisted and CAD-assisted detection and time efficiency of radiologists in reporting lung nodules on CT scans taken from patients with extra-thoracic malignancies using a Cloud-based system. MATERIALS AND

METHODS:

Three radiologists searched for pulmonary nodules in patients with extra-thoracic malignancy who underwent CT (slice thickness/spacing 2 mm/1.7 mm) between September 2015 and March 2016. All nodules detected by unassisted reading were measured and coordinates were uploaded on a cloud-based system. CAD marks were then reviewed by the same readers using the cloud-based interface. To establish the reference standard all nodules ≥ 3 mm detected by at least one radiologist were validated by two additional experienced radiologists in consensus. Reader detection rate and reporting time with and without CAD were compared. The study was approved by the local ethics committee. All patients signed written informed consent.

RESULTS:

The series included 225 patients (age range 21-90 years, mean 62 years), including 75 patients having at least one nodule, for a total of 215 nodules. Stand-alone CAD sensitivity for lesions ≥ 3 mm was 85% (183/215, 95% CI 82-91); mean false-positive rate per scan was 3.8. Sensitivity across readers in detecting lesions ≥ 3 mm was statistically higher using CAD 65% (95% CI 61-69) versus 88% (95% CI 86-91, p<0.01). Reading time increased by 11% using CAD (296 s vs. 329 s; p<0.05).

CONCLUSION:

In patients with extra-thoracic malignancies, CAD-assisted reading improves detection of ≥ 3-mm lung nodules on CT, slightly increasing reading time. KEY POINTS • CAD-assisted reading improves the detection of lung nodules compared with unassisted reading on CT scans of patients with primary extra-thoracic tumour, slightly increasing reading time. • Cloud-based CAD systems may represent a cost-effective solution since CAD results can be reviewed while a separated cloud back-end is taking care of computations. • Early identification of lung nodules by CAD-assisted interpretation of CT scans in patients with extra-thoracic primary tumours is of paramount importance as it could anticipate surgery and extend patient life expectancy.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Interpretação de Imagem Radiográfica Assistida por Computador / Nódulos Pulmonares Múltiplos / Computação em Nuvem / Neoplasias Pulmonares Tipo de estudo: Diagnostic_studies / Observational_studies / Prognostic_studies / Screening_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Interpretação de Imagem Radiográfica Assistida por Computador / Nódulos Pulmonares Múltiplos / Computação em Nuvem / Neoplasias Pulmonares Tipo de estudo: Diagnostic_studies / Observational_studies / Prognostic_studies / Screening_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2019 Tipo de documento: Article