Evaluation of pulmonary artery stenosis in congenital heart disease patients using functional diagnostic parameters: An in vitro study.
D'Souza, Gavin A; Banerjee, Rupak K; Taylor, Michael D.
Afiliação
  • D'Souza GA; Department of Mechanical and Materials Engineering, University of Cincinnati, Cincinnati, OH, USA.
  • Banerjee RK; Department of Mechanical and Materials Engineering, University of Cincinnati, Cincinnati, OH, USA. Electronic address: Rupak.Banerjee@uc.edu.
  • Taylor MD; The Heart Institute, Cincinnati Children's Hospital Medical Center, Cincinnati, OH, USA. Electronic address: michaeldtaylor@mac.com.
J Biomech ; 81: 58-67, 2018 11 16.
Article em En | MEDLINE | ID: mdl-30293825
ABSTRACT
Congenital pulmonary artery (PA) stenosis is often associated with abnormal PA hemodynamics including increased pressure drop (Δp) and reduced asymmetric flow (Q), which may result in right ventricular dysfunction. We propose functional diagnostic parameters, pressure drop coefficient (CDP), energy loss (Eloss), and normalized energy loss (E¯loss) to characterize pulmonary hemodynamics, and evaluate their efficacy in delineating stenosis severity using in vitro experiments. Subject-specific test sections including the main PA (MPA) bifurcating into left and right PAs (LPA, RPA) with a discrete LPA stenosis were manufactured from cross-sectional imaging and 3D printing. Three clinically-relevant stenosis severities, 90% area stenosis (AS), 80% AS, and 70% AS, were evaluated at different cardiac outputs (COs). A benchtop flow loop simulating pulmonary hemodynamics was used to measure Q and Δp within the test sections. The experimental Δp-Q characteristics along with clinical data were used to obtain pathophysiologic conditions and compute the diagnostic parameters. The pathophysiologic QLPA decreased as the stenosis severity increased at a fixed CO. CDPLPA, Eloss,LPA (absolute), and E¯loss,LPA (absolute) increased with an increase in LPA stenosis severity at a fixed CO. Importantly, CDPLPA and E¯loss,LPA had reduced variability with CO, and distinct values for each LPA stenosis severity. Under variable CO, a) CDPLPA values were 14.5-21.0 (70% AS), 60.7- 2.2 (80% AS), ≥ 261.6 (90% AS), and b) E¯loss,LPA values (in mJ per QLPA) were -501.9 to -1023.8 (70% AS), -1247.6 to -1773.0 (80% AS), -1934.5 (90% AS). Hence, CDPLPA and E¯loss,LPA are expected to assess the true functional severity of PA stenosis.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Estenose de Artéria Pulmonar / Cardiopatias Congênitas Tipo de estudo: Diagnostic_studies Limite: Female / Humans / Infant Idioma: En Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Estenose de Artéria Pulmonar / Cardiopatias Congênitas Tipo de estudo: Diagnostic_studies Limite: Female / Humans / Infant Idioma: En Ano de publicação: 2018 Tipo de documento: Article