Improving the Spatial Alignment in PET/CT Using Amplitude-Based Respiration-Gated PET and Patient-Specific Breathing-Instructed CT.
van der Vos, Charlotte S; Meeuwis, Antoi P W; Grootjans, Willem; Geus-Oei, Lioe-Fee de; Visser, Eric P.
Afiliação
  • van der Vos CS; Department of Radiology and Nuclear Medicine, Radboud University Medical Center, Nijmegen, The Netherlands charlotte.vandervos@radboudumc.nl.
  • Meeuwis APW; University of Twente, Enschede, The Netherlands; and.
  • Grootjans W; Department of Radiology and Nuclear Medicine, Radboud University Medical Center, Nijmegen, The Netherlands.
  • Geus-Oei LF; Department of Radiology, Leiden University Medical Center, Leiden, The Netherlands.
  • Visser EP; University of Twente, Enschede, The Netherlands; and.
J Nucl Med Technol ; 47(2): 154-159, 2019 Jun.
Article em En | MEDLINE | ID: mdl-30413602
ABSTRACT
Appropriate attenuation correction is important for accurate quantification of SUVs in PET. Patient respiratory motion can introduce a spatial mismatch between respiration-gated PET and CT, reducing quantitative accuracy. In this study, the effect of a patient-specific breathing-instructed CT protocol on the spatial alignment between CT and amplitude-based optimal respiration-gated PET images was investigated.

Methods:

18F-FDG PET/CT imaging was performed on 20 patients. In addition to the standard low-dose free-breathing CT, breath-hold CT was performed. The amplitude limits of the respiration-gated PET were used to instruct patients to hold their breath during CT acquisition at a similar amplitude level. Spatial mismatch was quantified using the position differences between the lung-liver transition in PET and CT images, the distance between PET and CT lesions' centroids, and the amount of overlap as indicated by the Jaccard similarity coefficient. Furthermore, the effect on attenuation correction was quantified by measuring SUVs, metabolic tumor volume, and total lesion glycolysis (TLG) of lung lesions.

Results:

All patients found the breathing instructions feasible; however, 4 patients had trouble complying with the instructions. In total, 18 patients were included. The average distance between the lung-liver transition between PET and CT was significantly reduced for breath-hold CT (1.7 ± 2.1 mm), compared with standard CT (5.6 ± 7.3 mm) (P = 0.049). Furthermore, the mean distance between the lesions' centroids on PET and CT was significantly smaller for breath-hold CT (3.6 ± 2.0 mm) than for standard CT (5.5 ± 6.5 mm) (P = 0.040). Quantification of lung lesion SUV was significantly affected, with a higher SUVmean when breath-hold CT (6.3 ± 3.9 g/cm3) was used for image reconstruction than for standard CT (6.1 ± 3.8 g/cm3) (P = 0.044). Though metabolic tumor volume was not significantly different, TLG reached statistical significance.

Conclusion:

Optimal respiration-gated PET in combination with patient-specific breathing-instructed CT results in an improved alignment between PET and CT images and shows an increased SUVmean and TLG. Even though the effects are small, a more accurate SUV and TLG determination is of importance for a more stable PET quantification, which is relevant for radiotherapy planning and therapy response monitoring.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Respiração / Técnicas de Imagem de Sincronização Respiratória / Tomografia por Emissão de Pósitrons combinada à Tomografia Computadorizada Tipo de estudo: Guideline Limite: Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Respiração / Técnicas de Imagem de Sincronização Respiratória / Tomografia por Emissão de Pósitrons combinada à Tomografia Computadorizada Tipo de estudo: Guideline Limite: Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2019 Tipo de documento: Article