Fatigue and microgap behaviour of a three-unit implant-fixed dental prosthesis combining conventional and dynamic abutments.
Rocha Ferreira, José; Song, Nannan; Machado, Margarida; Machado, Luís; Correia, André; Reis, Ana.
Afiliação
  • Rocha Ferreira J; a Advanced Manufacturing Technologies Department, Institute of Mechanical Engineering and Industrial Management (INEGI) , Oporto , Portugal.
  • Song N; a Advanced Manufacturing Technologies Department, Institute of Mechanical Engineering and Industrial Management (INEGI) , Oporto , Portugal.
  • Machado M; a Advanced Manufacturing Technologies Department, Institute of Mechanical Engineering and Industrial Management (INEGI) , Oporto , Portugal.
  • Machado L; b Centre of Mathematics , University of Minho , Guimarães , Portugal.
  • Correia A; c Health Interdisciplinary Research Centre of the Institute of Health Sciences of Viseu , Portuguese Catholic University , Portugal.
  • Reis A; a Advanced Manufacturing Technologies Department, Institute of Mechanical Engineering and Industrial Management (INEGI) , Oporto , Portugal.
J Med Eng Technol ; 42(5): 397-405, 2018 Jul.
Article em En | MEDLINE | ID: mdl-30422735
ABSTRACT
This is an in vitro study composed by a fatigue test followed by an optical microscopy analysis. Dynamic abutments concept, recently introduced on screw-retained implant dental prosthesis, consists on the screw channel customisation according to the individual needs of each rehabilitation. Geometry and tightening torque differences advise the assessment of their mechanical performance. Clarify whether the combination of dynamic and conventional abutments in a three-unit implant-fixed prosthesis has detrimental effects either on the mechanical performance under cyclic loading or on the implant-abutment microgap dimensions. The fatigue test was performed in agreement with the ISO standard 14801. Then on the samples that resisted 5 million cycles, the implant-abutment microgap was measured on dynamic and conventional abutments using optical microscopy. Two unloaded samples were used as control group. The samples supported a load of 1050 N. The implant-abutment microgap measurement did not show statistically significant differences (p = .086) between loaded and unloaded groups, but the loaded conventional abutments showed a significant lower implant-abutment microgap (p = .05) than the loaded dynamic abutments. The combination of conventional and dynamic abutments do not seem to produce a decrease in fatigue resistance to a level below the mastication forces or an increase in the joint dimensions.