Variation of degree of stenosis quantification using different energy level with dual energy CT scanner.
Saba, Luca; Argioas, Giovanni Maria; Lucatelli, Pierleone; Lavra, Francesco; Suri, Jasjit S; Wintermark, Max.
Afiliação
  • Saba L; Department of Radiology, Azienda Ospedaliero Universitaria (A.O.U.) di Cagliari, Polo di Monserrato s.s. 554 Monserrato, 09045, Cagliari, Italy. lucasaba@tiscali.it.
  • Argioas GM; Department of Radiology, AOB di Cagliari, 09045, Cagliari, Italy.
  • Lucatelli P; Department of Radiological, Oncological and Anatomopathological Sciences-Radiology, 'Sapienza' University of Rome, Rome, Italy.
  • Lavra F; Department of Radiology, Azienda Ospedaliero Universitaria (A.O.U.) di Cagliari, Polo di Monserrato s.s. 554 Monserrato, 09045, Cagliari, Italy.
  • Suri JS; Diagnostic and Monitoring Division, AtheroPoint™ LLC, Roseville, CA, 95661, USA.
  • Wintermark M; Department of Electrical Engineering, University of Idaho (Affl.), Moscow, ID, USA.
Neuroradiology ; 61(3): 285-291, 2019 Mar.
Article em En | MEDLINE | ID: mdl-30554271
PURPOSE: To investigate the variation in the quantification of the carotid degree of stenosis (DoS) with a dual energy computed tomography (CT), using different energy levels during the image reconstruction. METHODS: In this retrospective study, 53 subjects (37 males; mean age 67 ± 11 years; age range 47-83 years) studied with a multi-energy CT scanner were included. Datasets were reconstructed on a dedicated workstation and from the CT raw data multiple datasets were generated at the following monochromatic energy levels: 66, 70, 77, and 86 kilo-electronvolt (keV). Two radiologists independently performed all measurements for quantification of the degree of stenosis. Wilcoxon test was used to test the differences between the Hounsifield unit (HU) values in the plaques at different keV. RESULTS: The Wilcoxon analysis showed a statistically significant difference (p = 0.001) in the DoS assessment among the different keVs selected. The Bland-Altman analysis showed that the DoS difference had a linear relation with the keV difference (the bigger is the difference in keV, the bigger is the variation in DoS) and that for different keVs, the difference in DoS is reduced with its increase. CONCLUSION: A standardization in the use of the energy level during the image reconstruction should be considered.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Imagem Radiográfica a Partir de Emissão de Duplo Fóton / Estenose das Carótidas Tipo de estudo: Observational_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Imagem Radiográfica a Partir de Emissão de Duplo Fóton / Estenose das Carótidas Tipo de estudo: Observational_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2019 Tipo de documento: Article