Bulk Ti nitride prepared from rutile TiO2 for its application as stimulation electrode in neuroscience.
Canillas, M; Moreno, B; Carballo-Vila, M; Jurado, J R; Chinarro, E.
Afiliação
  • Canillas M; Instituto de Cerámica y Vidrio (ICV), Consejo Superior de Investigaciones Científicas (CSIC), Madrid, Spain. Electronic address: mcanillas@icv.csic.es.
  • Moreno B; Instituto de Cerámica y Vidrio (ICV), Consejo Superior de Investigaciones Científicas (CSIC), Madrid, Spain; Hospital Nacional de Parapléjicos, Toledo, Spain.
  • Carballo-Vila M; Hospital Nacional de Parapléjicos, Toledo, Spain.
  • Jurado JR; Instituto de Cerámica y Vidrio (ICV), Consejo Superior de Investigaciones Científicas (CSIC), Madrid, Spain.
  • Chinarro E; Instituto de Cerámica y Vidrio (ICV), Consejo Superior de Investigaciones Científicas (CSIC), Madrid, Spain.
Mater Sci Eng C Mater Biol Appl ; 96: 295-301, 2019 Mar.
Article em En | MEDLINE | ID: mdl-30606535
ABSTRACT
Bulk titanium nitride (TiN) was synthesized by nitridation of TiO2 rutile substrates. TiN pellets were successfully achieved at 1100 °C in ammonia stream; these materials were characterized by the evaluation of their microstructure, surface, chemical composition and electrical and electrochemical properties, concluding that the synthesis promotes the creation of a TiNxOy surface, which shows high metallic conductivity (close to 102 S/cm) and a microstructure with micro- and nano-features. Electrochemical studies reveal high storage capacities which are delivered through an injection mechanism that involves the double charge layer and EIS show a high capacitive contribution to the mechanism. Neuron cell cultures assessed the biocompatibility of the sample prepared and put forward this material as a promising candidate for implantable stimulation electrode in neuroscience.