Assessing zinc from a nail clipping using mono-energetic portable X-ray fluorescence.
Fleming, David E B; Crook, Samantha L; Evans, Colby T.
Afiliação
  • Fleming DEB; Physics Department, Mount Allison University, Sackville, NB, Canada. Electronic address: dfleming@mta.ca.
  • Crook SL; Physics Department, Mount Allison University, Sackville, NB, Canada.
  • Evans CT; Physics Department, Mount Allison University, Sackville, NB, Canada.
Appl Radiat Isot ; 145: 170-175, 2019 Mar.
Article em En | MEDLINE | ID: mdl-30639633
ABSTRACT
A mono-energetic X-ray beam from a portable X-ray fluorescence device was used to excite characteristic X-rays from zinc in a series of nail clipping phantoms. Twenty nail clipping phantoms having equal zinc concentrations of ~40 µg/g, but with different physical characteristics, were measured individually for 300 s using a small diameter (~1 mm) X-ray beam. Energy spectra obtained from the measurements were analyzed using PyMca software. Zinc signal size varied widely between the different clippings, with a relative standard deviation of 41% observed in the combined signal from zinc Kα and Kß characteristic X-rays. Three different normalization approaches were introduced to account for variation in the amounts of sample interrogated by the X-ray beam. All three approaches produced similar results, and successfully reduced the relative standard deviation to between 12% and 13%. A clear trend was still observed, however, between the normalized zinc signal and the thickness of clipping measured. To account for this effect, normalized signals were adjusted to calculate "thickness-corrected" values. The relative standard deviation of these thickness-corrected values was 6.2%. Reproducibility of measurement from individual clippings was excellent, with relative standard deviations on the order of 1%, with or without normalization. Overall, this new method of measuring zinc in nail shows promise for the assessment of zinc status in humans using a portable device. The method is sensitive, rapid, and requires only a single nail clipping.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Espectrometria por Raios X / Zinco / Unhas Tipo de estudo: Evaluation_studies Limite: Humans Idioma: En Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Espectrometria por Raios X / Zinco / Unhas Tipo de estudo: Evaluation_studies Limite: Humans Idioma: En Ano de publicação: 2019 Tipo de documento: Article