Quantitative Immunoblotting of Cell Lines as a Standard to Validate Immunofluorescence for Quantifying Biomarker Proteins in Routine Tissue Samples.
J Vis Exp
; (143)2019 01 07.
Article
em En
| MEDLINE
| ID: mdl-30663702
Quantification of proteins of interest in formalin-fixed, paraffin-embedded (FFPE) tissue samples is important in clinical and research applications. An optimal method of quantification is accurate, has a broad linear dynamic range and maintains the structural integrity of the sample to allow for identification of individual cell types. Current methods such as immunohistochemistry (IHC), mass spectrometry, and immunoblotting each fail to meet these stipulations due to their categorical nature or need to homogenize the sample. As an alternative method, we propose the use of immunofluorescence (IF) and image analysis to determine the relative abundance of a protein of interest in FFPE tissues. Herein we demonstrate that this method is easily optimized, yields a wide dynamic range, and is linearly quantifiable as compared to the gold standard of quantitative immunoblotting. Furthermore, this method permits the maintenance of the structural integrity of the sample and allows for the distinction of various cell types, which may be crucial in diagnostic applications. Overall, this is a robust method for the relative quantification of proteins in FFPE samples and can be easily adapted to suit clinical or research needs.
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Biomarcadores
/
Immunoblotting
/
Proteínas
/
Imunofluorescência
Tipo de estudo:
Prognostic_studies
Limite:
Humans
Idioma:
En
Ano de publicação:
2019
Tipo de documento:
Article