Your browser doesn't support javascript.
loading
Stapes Prosthesis Length: One Size Fits All?
Sudhoff, Holger; Gehl, Hans Björn; Boga, Ercan; Müller, Stefan; Wilms, Katharina; Mutze, Sven; Todt, Ingo.
Afiliação
  • Sudhoff H; Department of Otolaryngology, Head and Neck Surgery, Klinikum Bielefeld, Academic Hospital: Ruhr Universität Bochum, Bielefeld, Germany.
  • Gehl HB; Department of Radiology, Klinikum Bielefeld, Bielefeld, Germany.
  • Boga E; Department of Radiology, Klinikum Bielefeld, Bielefeld, Germany.
  • Müller S; Department of Otolaryngology, Head and Neck Surgery, Klinikum Bielefeld, Academic Hospital: Ruhr Universität Bochum, Bielefeld, Germany.
  • Wilms K; Department of Otolaryngology, Head and Neck Surgery, Unfallkrankenhaus Berlin, Berlin, Germany.
  • Mutze S; Department of Radiology, Unfallkrankenhaus Berlin, Berlin, Germany.
  • Todt I; Department of Otolaryngology, Head and Neck Surgery, Klinikum Bielefeld, Academic Hospital: Ruhr Universität Bochum, Bielefeld, Germany, todt@gmx.net.
Audiol Neurootol ; 24(1): 1-7, 2019.
Article em En | MEDLINE | ID: mdl-30783032
ABSTRACT

BACKGROUND:

The insertion of the stapes piston into the vestibule provides the physical basis for a successful stapedotomy. In routine clinical practice, two different ways to handle prosthesis length are performed (1) an individualized measurement of the stapes prosthesis length or (2) a standard prosthesis length for all cases.

OBJECTIVE:

The objective of this study was to compare both ways of handling prosthesis length and the effect of these methods on insertional prosthesis depth. MATERIAL AND

METHOD:

We retrospectively evaluated 39 patients after performing a stapedotomy for radiologically estimated vestibular stapes prosthesis insertion depth. The individual measured length data were hypothetically changed to a standard length of 4.75, 5, 5.25, and 5.5 mm, and the insertion depths were compared.

RESULTS:

The individually measured prosthesis lengths led to an insertion depth between 0.2 and 1.6 mm (mean 0.74 mm). The ratio of insertion depth/vestibular depth was between 8 and 59.1% (mean 26.6%). The different assumed standard lengths led to different rates of the vestibulum positions and possible bony contacts at the vestibulum floor.

CONCLUSION:

The individual measurement led to a zero rate of the vestibulum positions of stapes prosthesis pistons with a low insertion depth/vestibular depth ratio.