Patient-specific SPECT imaging protocols to standardize image noise.
Cuddy-Walsh, Sarah G; Clackdoyle, Duncan C; Renaud, Jennifer M; Wells, R Glenn.
Afiliação
  • Cuddy-Walsh SG; Department of Physics, Carleton University, Ottawa, ON, Canada. swalsh@ottawaheart.ca.
  • Clackdoyle DC; Division of Cardiology, Cardiac Imaging, University of Ottawa Heart Institute, H2258 - 40 Ruskin Street, Ottawa, ON, K1Y 4W7, Canada. swalsh@ottawaheart.ca.
  • Renaud JM; Division of Cardiology, Cardiac Imaging, University of Ottawa Heart Institute, H2258 - 40 Ruskin Street, Ottawa, ON, K1Y 4W7, Canada.
  • Wells RG; Division of Cardiology, Cardiac Imaging, University of Ottawa Heart Institute, H2258 - 40 Ruskin Street, Ottawa, ON, K1Y 4W7, Canada.
J Nucl Cardiol ; 28(1): 225-233, 2021 02.
Article em En | MEDLINE | ID: mdl-30834500
BACKGROUND: In addition to acquired photon counts, image noise depends on the image reconstruction algorithm. This work develops patient-specific activity or acquisition time protocols to standardize the average noise in a reconstructed image for different patients, cameras, and reconstruction algorithms. METHODS: Image noise was calculated for images from 43 patients acquired on both a conventional and a multiple-pinhole cardiac SPECT camera. Functions were found to relate image noise to radiotracer activity, scan time, and body mass and were validated by normalizing the image noise in a test set of 58 patients. RESULTS: There was a 3.6-fold difference in photon sensitivity between the two cameras but a 16-fold difference in activity-scan time was necessary to match the noise levels. Image noise doubled from 45 to 128 kg for the conventional camera (12.8 minutes) and tripled for the multiple-pinhole camera (5 minutes) for 350 MBq (9.5 mCi) 99mTc-tetrofosmin. It was 16.3% and 6.1% respectively for an average sized patient. CONCLUSIONS: A linear scaling of activity with respect to the patient weight normalizes image noise but the scaling factors depend on the choice of camera and image reconstruction parameters. Therefore, equivalent numbers of acquired photon counts are not sufficient to guarantee equivalent image noise.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Câmaras gama / Tomografia Computadorizada de Emissão de Fóton Único / Imagem de Perfusão do Miocárdio / Coração Tipo de estudo: Diagnostic_studies / Guideline / Observational_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Câmaras gama / Tomografia Computadorizada de Emissão de Fóton Único / Imagem de Perfusão do Miocárdio / Coração Tipo de estudo: Diagnostic_studies / Guideline / Observational_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2021 Tipo de documento: Article