Divided-volume matching technique for volume displacement estimation of patient positioning in radiation therapy.
Tashiro, Mutsumi; Kubota, Yoshiki; Torikoshi, Masami; Ohno, Tatsuya; Nakano, Takashi.
Afiliação
  • Tashiro M; Gunma University Initiative for Advanced Research, 3-39-22 Showa-Machi, Maebashi, Gunma 371-8511, Japan; Gunma University Heavy Ion Medical Center, 3-39-22 Showa-Machi, Maebashi, Gunma 371-8511, Japan. Electronic address: tashiro@gunma-u.ac.jp.
  • Kubota Y; Gunma University Heavy Ion Medical Center, 3-39-22 Showa-Machi, Maebashi, Gunma 371-8511, Japan.
  • Torikoshi M; Gunma University Heavy Ion Medical Center, 3-39-22 Showa-Machi, Maebashi, Gunma 371-8511, Japan.
  • Ohno T; Gunma University Heavy Ion Medical Center, 3-39-22 Showa-Machi, Maebashi, Gunma 371-8511, Japan.
  • Nakano T; Gunma University Initiative for Advanced Research, 3-39-22 Showa-Machi, Maebashi, Gunma 371-8511, Japan; Gunma University Heavy Ion Medical Center, 3-39-22 Showa-Machi, Maebashi, Gunma 371-8511, Japan.
Phys Med ; 62: 1-12, 2019 Jun.
Article em En | MEDLINE | ID: mdl-31153388
PURPOSE: We propose the Divided-Volume Matching (DVM) technique to visualize and estimate three-dimensional (3D) displacements of internal structures to enable more accurate patient positioning for radiation therapy. METHODS: A CT volume is divided into a volume of interest (VOI) and a base volume (BV); 2D-3D matching is achieved using digital radiography (DR) images and digitally reconstructed radiographs (DRRs), where the DRRs are iteratively generated by changing the 3D positions and rotation angles of the separate volumes independently to identify the best match with the DR images. We demonstrate this technique with two phantom and two clinical cases. RESULTS: 3D displacements of the VOIs could be estimated independently and simultaneously with those of the BVs, with accuracies comparable to those of the conventional 2D-3D matching. The proposed technique yielded more suitable matching results when internal displacements occurred in the regions of interest (ROIs). The best matches were found when the ROI was confined to the focused structure, initial displacement values were coarsely adjusted, one volume was matched while the other was fixed, or any combination thereof. CONCLUSIONS: The proposed technique can be used effectively for independent displacement estimations of VOIs and BVs for patient positioning in radiation therapy.