Your browser doesn't support javascript.
loading
Wide-volume versus helical acquisition in unenhanced chest CT: prospective intra-patient comparison of diagnostic accuracy and radiation dose in an ultra-low-dose setting.
Meyer, Elsa; Labani, Aissam; Schaeffer, Mickaël; Jeung, Mi-Young; Ludes, Claire; Meyer, Alain; Roy, Catherine; Leyendecker, Pierre; Ohana, Mickaël.
Afiliação
  • Meyer E; Radiology Department, Nouvel Hôpital Civil, 1 place de l'Hôpital, 67000, Strasbourg, France.
  • Labani A; Radiology Department, Nouvel Hôpital Civil, 1 place de l'Hôpital, 67000, Strasbourg, France.
  • Schaeffer M; Radiology Department, Nouvel Hôpital Civil, 1 place de l'Hôpital, 67000, Strasbourg, France.
  • Jeung MY; Radiology Department, Nouvel Hôpital Civil, 1 place de l'Hôpital, 67000, Strasbourg, France.
  • Ludes C; Radiology Department, Nouvel Hôpital Civil, 1 place de l'Hôpital, 67000, Strasbourg, France.
  • Meyer A; Physiology Department, Nouvel Hôpital Civil, 1 place de l'Hôpital, 67000, Strasbourg, France.
  • Roy C; Radiology Department, Nouvel Hôpital Civil, 1 place de l'Hôpital, 67000, Strasbourg, France.
  • Leyendecker P; Radiology Department, Nouvel Hôpital Civil, 1 place de l'Hôpital, 67000, Strasbourg, France.
  • Ohana M; Radiology Department, Nouvel Hôpital Civil, 1 place de l'Hôpital, 67000, Strasbourg, France. mickael.ohana@gmail.com.
Eur Radiol ; 29(12): 6858-6866, 2019 Dec.
Article em En | MEDLINE | ID: mdl-31175414
ABSTRACT

OBJECTIVES:

Diagnostic performance and potential radiation dose reduction of wide-area detector CT sequential acquisition ("wide-volume" acquisition (WV)) in unenhanced chest examination are unknown. This study aims to assess the image quality, the diagnostic performance, and the radiation dose reduction of WV mode compared with the classical helical acquisition for lung parenchyma analysis in an ultra-low-dose (ULD) protocol.

METHODS:

After Institutional Review Board Approval and written informed consent, 64 patients (72% men; 67.6 ± 9.7 years old; BMI 26.1 ± 5.3 kg/m2) referred for a clinically indicated unenhanced chest CT were prospectively included. All patients underwent, in addition to a standard helical acquisition (120 kV, automatic tube current modulation), two ULD acquisitions (135 kV, fixed tube current at 10 mA) one in helical mode and one in WV mode. Image noise, subjective image quality (5-level Likert scale), and diagnostic performance for the detection of 9 predetermined parenchymal abnormalities were assessed by two radiologists and compared using the chi-square or Fisher non-parametric tests.

RESULTS:

Subjective image quality (4.2 ± 0.7 versus 4.2 ± 0.8, p = 0.56), image noise (41.7 ± 8 versus 40.9 ± 8.7, p = 0.3), and diagnostic performance were equivalent between ULD WV and ULD helical. Radiation dose was significantly lower for the ULD WV acquisition (mean dose-length product 14.1 ± 1.3 mGy cm versus 15.8 ± 1.3, p < 0.0001).

CONCLUSION:

An additional 11% dose reduction is achieved with the WV mode in ULD chest CT with fixed tube current, with equivalent image quality and diagnostic performance when compared with the helical acquisition. KEY POINTS • Image quality and diagnostic performance of ultra-low-dose unenhanced chest CT are identical between wide-volume mode and the reference helical acquisition. • Wide-volume mode allows an additional radiation dose reduction of 11% (mean dose-length product 14.1 ± 1.3 mGy cm versus 15.8 ± 1.3, p < 0.0001).
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Doses de Radiação / Radiografia Torácica / Interpretação de Imagem Radiográfica Assistida por Computador / Tomografia Computadorizada por Raios X Tipo de estudo: Diagnostic_studies / Guideline / Observational_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Doses de Radiação / Radiografia Torácica / Interpretação de Imagem Radiográfica Assistida por Computador / Tomografia Computadorizada por Raios X Tipo de estudo: Diagnostic_studies / Guideline / Observational_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2019 Tipo de documento: Article