Your browser doesn't support javascript.
loading
Percutaneous Lung Tumor Biopsy Under CBCT Guidance with PET-CT Fusion Imaging: Preliminary Experience.
Floridi, C; Carnevale, A; Fumarola, E M; Schampaert, S; Fontana, F; De Palma, D; Del Sole, A; Giganti, M; Carrafiello, G.
Afiliação
  • Floridi C; Department of Health Sciences, Diagnostic and Interventional Radiology, San Paolo Hospital, University of Milan, Milan, Italy. chiara.floridi@gmail.com.
  • Carnevale A; Diagnostic and Interventional Radiology Department, San Paolo Hospital, Via Di Rudinì, 20121, Milan, Italy. chiara.floridi@gmail.com.
  • Fumarola EM; University Radiology Unit, Department of Radiology, University Hospital of Ferrara, Ferrara, Italy.
  • Schampaert S; Department of Health Sciences, Diagnostic and Interventional Radiology, San Paolo Hospital, University of Milan, Milan, Italy.
  • Fontana F; Image-Guided Therapy Systems, Philips Healthcare, Best, The Netherlands.
  • De Palma D; Radiology Department, Ospedale di Circolo e Fondazione Macchi, Varese, Italy.
  • Del Sole A; Department of Nuclear Medicine, Ospedale di Circolo e Fondazione Macchi, Varese, Italy.
  • Giganti M; Nuclear Medicine Unit, ASST Santi Paolo e Carlo, University of Milan, Milan, Italy.
  • Carrafiello G; Department of Morphology, Surgery and Experimental Medicine, Section of Radiology, University of Ferrara, Ferrara, Italy.
Cardiovasc Intervent Radiol ; 42(11): 1644-1648, 2019 Nov.
Article em En | MEDLINE | ID: mdl-31218410
ABSTRACT

PURPOSE:

The aim of this study is to evaluate the feasibility of percutaneous lung tumor biopsy under cone beam-computed tomography (CBCT) with PET-CT imaging fusion. MATERIALS AND

METHODS:

Eleven patients (four women and seven men) underwent C-arm CBCT lung biopsy with PET-CT fusion imaging. A preprocedural PET-CT scan was manually fused with procedural CBCT based on anatomical landmarks; using real-time fluoroscopy, the coregistered PET-CT and CBCT images were overlaid to guide the needle trajectory. Technical success, accuracy, sensibility and specificity were evaluated. Mean total procedure time and time required for image elaboration were recorded.

RESULTS:

Technical success, diagnostic accuracy, sensitivity and specificity were 100%. The mean procedure time was 38 min. The average time of PET-CT/CBCT image fusion elaboration was 3.53 min for planning and 3.42 min for needle positioning check.

CONCLUSION:

CBCT-guided percutaneous lung biopsy with PET-CT fusion imaging is a feasible and effective procedure, with the potential to further improve diagnostic yield by targeting the most metabolically active portion of a lesion, whether it is morphologically altered or normal.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Radiografia Intervencionista / Tomografia Computadorizada de Feixe Cônico / Tomografia por Emissão de Pósitrons combinada à Tomografia Computadorizada / Neoplasias Pulmonares Tipo de estudo: Diagnostic_studies / Guideline / Observational_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Radiografia Intervencionista / Tomografia Computadorizada de Feixe Cônico / Tomografia por Emissão de Pósitrons combinada à Tomografia Computadorizada / Neoplasias Pulmonares Tipo de estudo: Diagnostic_studies / Guideline / Observational_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2019 Tipo de documento: Article