Your browser doesn't support javascript.
loading
Evaluation of an immunoturbidimetric microalbuminuria assay.
Paloheimo, L; Pajari-Backas, M; Pitkänen, E; Melamies, L; Rissanen, R.
Afiliação
  • Paloheimo L; United Laboratories Ltd, Helsinki, Finland.
J Clin Chem Clin Biochem ; 25(12): 889-92, 1987 Dec.
Article em En | MEDLINE | ID: mdl-3127530
We have evaluated an immunoturbidimetric method for the estimation of urinary albumin. The method, besides being easy to perform and cost-effective, was sensitive enough to detect an even slightly increased albumin excretion (detection limit 5 mg/l). Within-run reproducibility was 1.8 and 2.1%, and between-run reproducibility 2.9 and 4.3% in samples containing 16.1-17.8 mg/l and 50.6-54.0 mg/l of albumin, respectively. The recovery of albumin added to the samples was 98.6-106.6%. Results obtained by this method correlated well with the results obtained by radial immunodiffusion (r = 0.980, n = 44) and radioimmunoassay (r = 0.982, n = 41). The immunoturbidimetric method can be easily adapted for several clinical chemistry analysers.
Assuntos
Buscar no Google
Eixos temáticos: Pesquisa_clinica Base de dados: MEDLINE Assunto principal: Técnicas Imunológicas / Albuminúria / Nefelometria e Turbidimetria Tipo de estudo: Diagnostic_studies / Evaluation_studies Limite: Humans Idioma: En Ano de publicação: 1987 Tipo de documento: Article
Buscar no Google