Your browser doesn't support javascript.
loading
Corneal Endothelial Cell Loss and Morphometric Changes 5 Years after Phacoemulsification with or without CyPass Micro-Stent.
Lass, Jonathan H; Benetz, Beth Ann; He, Jonathan; Hamilton, Cody; Von Tress, Mark; Dickerson, Jaime; Lane, Stephen.
Afiliação
  • Lass JH; Cornea Image Analysis Reading Center, University Hospitals Eye Institute, Department of Ophthalmology and Visual Sciences, Case Western Reserve University, Cleveland, Ohio, USA.
  • Benetz BA; Cornea Image Analysis Reading Center, University Hospitals Eye Institute, Department of Ophthalmology and Visual Sciences, Case Western Reserve University, Cleveland, Ohio, USA.
  • He J; Alcon Research, LLC, Fort Worth, Texas, USA.
  • Hamilton C; Alcon Research, LLC, Fort Worth, Texas, USA.
  • Von Tress M; Alcon Research, LLC, Fort Worth, Texas, USA.
  • Dickerson J; Alcon Research, LLC, Fort Worth, Texas, USA; North Texas Eye Research Institute, University of North Texas Health Science Center, Fort Worth, Texas, USA.
  • Lane S; Alcon Research, LLC, Fort Worth, Texas, USA; Department of Ophthalmology, University of Minnesota, Minneapolis, Minnesota, USA. Electronic address: stephen.lane@alcon.com.
Am J Ophthalmol ; 208: 211-218, 2019 12.
Article em En | MEDLINE | ID: mdl-31377278
ABSTRACT

PURPOSE:

To characterize long-term changes in corneal endothelial cells after phacoemulsification with or without supraciliary Micro-Stent (Alcon) implantation in eyes with open-angle glaucoma (OAG) and visually significant cataract.

DESIGN:

Three-year safety extension of a 2-year randomized clinical trial.

METHODS:

Patients from the multicenter Study of an Implantable Device for Lowering Intraocular Pressure in Glaucoma Patients Undergoing Cataract Surgery (COMPASS) trial who underwent Micro-Stent implantation plus phacoemulsification (n = 282) or phacoemulsification alone (n = 67) were analyzed post hoc. Specular microscopy was used to assess endothelial cell loss (ECL), including changes from baseline in endothelial cell density (ECD), coefficient of variation, and percentage of hexagonal cells.

RESULTS:

Preoperative ECDs in the microstent group (2,432.6 cells/mm2 [95% confidence interval [CI], 2,382.8-2,482.4 cells/mm2]) were similar to those in the control group (2,434.5 cells/mm2 [95% CI, 2,356.5-2,512.4 cells/mm2]) groups. ECL at months 48 and 60 was greater in the Micro-Stent group than in the control group. At month 60, the mean percent of changes in ECD was -20.4% (95% CI, -23.5% to -17.5%) in the Micro-Stent group and -10.1% (95% CI, -13.9% to -6.3%) in the control group. No statistically significant between-group changes from baseline in cellular morphology were observed. Nine adverse events were possibly related to ECL, including 3 eyes with transient focal corneal edema and 4 eyes that required Micro-Stent trimming due to protrusion.

CONCLUSIONS:

In eyes with OAG, ECL after phacoemulsification is acute and stabilizes after 3 months, whereas ECL after phacoemulsification plus Micro-Stent implantation proceeds for at least 5 years. Clinical findings associated with ECL in these eyes were uncommon (3.3% of implanted eyes), suggesting that ECL is generally a subclinical phenomenon.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Complicações Pós-Operatórias / Endotélio Corneano / Glaucoma de Ângulo Aberto / Facoemulsificação / Implantes para Drenagem de Glaucoma / Perda de Células Endoteliais da Córnea Tipo de estudo: Clinical_trials / Diagnostic_studies / Observational_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Complicações Pós-Operatórias / Endotélio Corneano / Glaucoma de Ângulo Aberto / Facoemulsificação / Implantes para Drenagem de Glaucoma / Perda de Células Endoteliais da Córnea Tipo de estudo: Clinical_trials / Diagnostic_studies / Observational_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2019 Tipo de documento: Article