Expanded feasibility of ferritin testing: stability of ferritin stored as whole blood and validation of plastic tubes.
Spencer, Bryan R; Brodsky, Jaye P; Holley, Gary C; Foster, Gregory A; Winton, Colleen; Stramer, Susan L.
Afiliação
  • Spencer BR; American Red Cross Scientific Affairs, Dedham, Massachusetts.
  • Brodsky JP; American Red Cross Scientific Affairs, Gaithersburg, Maryland.
  • Holley GC; Quality Analytics, Riverwoods, Illinois.
  • Foster GA; American Red Cross Scientific Affairs, Dedham, Massachusetts.
  • Winton C; American Red Cross Scientific Affairs, Gaithersburg, Maryland.
  • Stramer SL; American Red Cross Scientific Affairs, Dedham, Massachusetts.
Transfusion ; 59(11): 3424-3430, 2019 11.
Article em En | MEDLINE | ID: mdl-31503347
BACKGROUND: Ferritin testing is a recommended strategy to mitigate iron depletion in blood donors. A barrier for some testing platforms is a requirement to complete sample management and testing within a temporal window incompatible with the logistics of many blood collectors. The ability to delay separation of plasma/serum from red cells and subsequent testing would enhance the feasibility of ferritin testing on a broader scale. STUDY DESIGN AND METHODS: Thirty blood donors provided a research donation of 12, 4-mL sample tubes of whole blood. Six pairs of serum and K2 -EDTA-plasma tubes were centrifuged and samples tested in triplicate on day of collection and on each of the next 5 days following storage at 4°C. Comparison of ferritin values for serum versus K2 -EDTA-plasma at baseline was performed to validate plastic EDTA-containing tubes. Variation of ferritin values during storage was assessed for direction and strength of any detectable changes. RESULTS: Ferritin values were comparable between EDTA-plasma and serum, with baseline values from EDTA-plasma samples 7% lower on average than serum (p < 0.0001 by paired t-test). Variability over five storage days was within approved parameters in the manufacturer's instructions. Within-run precision averaged 2% to 3% for each test day and within-subject precision across all samples averaged less than 5% for both serum and EDTA-plasma. Repeated measures showed no difference in changes during storage by tube type or day of testing. CONCLUSION: These results support flexible testing procedures, expanding the opportunity for blood centers to adopt this measure for assessing donor iron status.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Preservação de Sangue / Coleta de Amostras Sanguíneas / Ferritinas Limite: Humans Idioma: En Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Preservação de Sangue / Coleta de Amostras Sanguíneas / Ferritinas Limite: Humans Idioma: En Ano de publicação: 2019 Tipo de documento: Article