Study on the RBE estimation for carbon beam scanning irradiation using a solid-state microdosimeter.
Han, Soorim; Yoo, Seung Hoon; Shin, Jae Ik; Kim, Eun Ho; Jung, Won-Gyun; Kim, Kum Bae; Matsumura, Akihiko; Kanai, Tatsuaki; Tran, Linh T; Chartier, Lachlan; James, Benjamin; Rosenfeld, Anatoly B.
Afiliação
  • Han S; Division of Heavy-Ion Clinical Research, Korea Institute of Radiological and Medical Sciences, Seoul, Korea.
  • Yoo SH; Division of Heavy-Ion Clinical Research, Korea Institute of Radiological and Medical Sciences, Seoul, Korea.
  • Shin JI; Division of Heavy-Ion Clinical Research, Korea Institute of Radiological and Medical Sciences, Seoul, Korea.
  • Kim EH; Division of Heavy-Ion Clinical Research, Korea Institute of Radiological and Medical Sciences, Seoul, Korea.
  • Jung WG; Division of Heavy-Ion Clinical Research, Korea Institute of Radiological and Medical Sciences, Seoul, Korea.
  • Kim KB; Department of Radiation Oncology, Korea Institute of Radiological and Medical Sciences, Seoul, Korea.
  • Matsumura A; Gunma Heavy Ion Medical Center, Gunma, Japan.
  • Kanai T; Gunma Heavy Ion Medical Center, Gunma, Japan.
  • Tran LT; Center for Medical Radiation Physics, University of Wollongong, Wollongong, Australia.
  • Chartier L; Center for Medical Radiation Physics, University of Wollongong, Wollongong, Australia.
  • James B; Center for Medical Radiation Physics, University of Wollongong, Wollongong, Australia.
  • Rosenfeld AB; Center for Medical Radiation Physics, University of Wollongong, Wollongong, Australia.
Med Phys ; 47(2): 363-370, 2020 Feb.
Article em En | MEDLINE | ID: mdl-31732963
PURPOSE: The purpose of this study was to study the field size effect on the estimated Relative Biological Effectiveness (RBE) for carbon scanning beam irradiation. METHODS: A silicon-on-insulator (SOI) microdosimeter system developed by the Centre for Medical Radiation Physics, University of Wollongong, Australia, was used for lineal-energy measurements (microdosimetric quantity). The RBE values were derived based on the modified microdosimetric kinetic model (MKM) at different depths in a water phantom in the scanning carbon beam for various scanned areas. RESULTS: Our study shows that the difference in RBE values derived from the SOI microdosimeter measurements with the MKM model and from the Treatment Planning System (TPS). The difference of the RBE values is within 6.5 % at the peak point of the spread-out Bragg Peak (SOBP) region. Compared to the spot-beam, RBE values obtained in the scanned-beam with a larger scanned area of 1.0 × 1.0 cm2 have better agreement with which estimated by the TPS. CONCLUSIONS: This study shows the possibility of using the SOI microdosimeter system as a quality assurance (QA) tool for RBE evaluation in carbon-pencil beam scanning radiotherapy.