Your browser doesn't support javascript.
loading
Assessing the Use of Incorrectly Scaled Optical Coherence Tomography Angiography Images in Peer-Reviewed Studies: A Systematic Review.
Llanas, Samantha; Linderman, Rachel E; Chen, Fred K; Carroll, Joseph.
Afiliação
  • Llanas S; School of Medicine, Medical College of Wisconsin, Milwaukee.
  • Linderman RE; Department of Cell Biology, Neurobiology & Anatomy, Medical College of Wisconsin, Milwaukee.
  • Chen FK; Centre for Ophthalmology and Visual Science (incorporating Lions Eye Institute), The University of Western Australia, Perth, Western Australia, Australia.
  • Carroll J; Department of Ophthalmology, Royal Perth Hospital, Perth, Western Australia, Australia.
JAMA Ophthalmol ; 138(1): 86-94, 2020 01 01.
Article em En | MEDLINE | ID: mdl-31774456
Importance: Individual differences in axial length affect the lateral magnification of in vivo retinal images and as a result can affect the accuracy of quantitative measurements made from these images. As measurements from optical coherence tomography angiography (OCTA) images are becoming increasingly used in the diagnosis and monitoring of a wide range of diseases, evaluating which studies use correctly scaled images is crucial to their interpretation. Objective: To perform a systematic literature review to assess the percentage of articles that report correcting the scale of their OCTA images for individual differences in retinal magnification. Evidence Review: A PubMed (MEDLINE) search was conducted for articles on OCTA retinal imaging published between June 1, 2015, and June 1, 2018. Initial results included 7552 articles. Initial exclusion criteria removed studies of animal models, as well as reviews, letters, replies, comments, and image-based or photographic essays. Articles not written in English and those that required purchase from non-English language websites were excluded. Articles that did not use OCTA for imaging the retina were also excluded. Remaining articles were reviewed in detail to assess whether the OCTA measurements required correct lateral scaling, and if so, whether axial length was reported or used to scale the images. We also determined the number of articles that mentioned the lack of correct lateral scaling as a limitation of the study. Findings: A total of 989 articles were included in the detailed review. Of these, 509 were determined to require correct image scaling for their analyses, but only 41 (8.0%) report measuring and using axial length to correct the lateral scale of their OCTA images. Furthermore, of the 468 articles that did not correctly scale their images, only 18 (3.8%) mentioned this as a limitation to their study. Conclusions and Relevance: These findings suggest that most peer-reviewed articles in PubMed that use quantitative OCTA measurements use incorrectly scaled images. This could call into question the conclusions of such studies and warrants consideration by OCTA manufacturers, physicians, authors, journal reviewers, and journal editors.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Angiofluoresceinografia / Interpretação de Imagem Assistida por Computador / Aumento da Imagem / Interpretação Estatística de Dados / Tomografia de Coerência Óptica Tipo de estudo: Prognostic_studies / Systematic_reviews Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Angiofluoresceinografia / Interpretação de Imagem Assistida por Computador / Aumento da Imagem / Interpretação Estatística de Dados / Tomografia de Coerência Óptica Tipo de estudo: Prognostic_studies / Systematic_reviews Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article