Utilizing Machine Learning for Image Quality Assessment for Reflectance Confocal Microscopy.
Kose, Kivanc; Bozkurt, Alican; Alessi-Fox, Christi; Brooks, Dana H; Dy, Jennifer G; Rajadhyaksha, Milind; Gill, Melissa.
Afiliação
  • Kose K; Dermatology Service, Memorial Sloan Kettering Cancer Center, New York, New York, USA. Electronic address: kosek@mskcc.org.
  • Bozkurt A; Electrical and Computer Engineering, Northeastern University, Boston, Massachusetts, USA.
  • Alessi-Fox C; Caliber Imaging and Diagnostics, Inc, Rochester, New York, USA.
  • Brooks DH; Electrical and Computer Engineering, Northeastern University, Boston, Massachusetts, USA.
  • Dy JG; Electrical and Computer Engineering, Northeastern University, Boston, Massachusetts, USA.
  • Rajadhyaksha M; Dermatology Service, Memorial Sloan Kettering Cancer Center, New York, New York, USA.
  • Gill M; Department of Pathology, SUNY Downstate Medical Center, Brooklyn, New York, USA; SkinMedical Research and Diagnostics, PLLC, Dobbs Ferry, New York, USA.
J Invest Dermatol ; 140(6): 1214-1222, 2020 06.
Article em En | MEDLINE | ID: mdl-31838127
ABSTRACT
In vivo reflectance confocal microscopy (RCM) enables clinicians to examine lesions' morphological and cytological information in epidermal and dermal layers while reducing the need for biopsies. As RCM is being adopted more widely, the workflow is expanding from real-time diagnosis at the bedside to include a capture, store, and forward model with image interpretation and diagnosis occurring offsite, similar to radiology. As the patient may no longer be present at the time of image interpretation, quality assurance is key during image acquisition. Herein, we introduce a quality assurance process by means of automatically quantifying diagnostically uninformative areas within the lesional area by using RCM and coregistered dermoscopy images together. We trained and validated a pixel-level segmentation model on 117 RCM mosaics collected by international collaborators. The model delineates diagnostically uninformative areas with 82% sensitivity and 93% specificity. We further tested the model on a separate set of 372 coregistered RCM-dermoscopic image pairs and illustrate how the results of the RCM-only model can be improved via a multimodal (RCM + dermoscopy) approach, which can help quantify the uninformative regions within the lesional area. Our data suggest that machine learning-based automatic quantification offers a feasible objective quality control measure for RCM imaging.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Pele / Dermatopatias / Processamento de Imagem Assistida por Computador / Dermoscopia / Aprendizado de Máquina Tipo de estudo: Clinical_trials / Diagnostic_studies / Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Pele / Dermatopatias / Processamento de Imagem Assistida por Computador / Dermoscopia / Aprendizado de Máquina Tipo de estudo: Clinical_trials / Diagnostic_studies / Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article