Aortic Root Pressurizing Device: Aortic Valve Evaluation During Cardioplegic Arrest.
Steindl, Johannes; Kirnbauer, Michael; Fierlbeck, Johann; Krieghofer, Michael; Hitzl, Wolfgang; Hinterbuchner, Lynne; Seitelberger, Rainald; Dinges, Christian.
Afiliação
  • Steindl J; Department of Cardiovascular & Endovascular Surgery, Paracelsus Medical University, Salzburg, Austria.
  • Kirnbauer M; Department of Anesthesiology, Perioperative Care and Intensive Care Medicine, Paracelsus Medical University, Salzburg, Austria.
  • Fierlbeck J; PMU Innovations GmbH, Salzburg, Austria.
  • Krieghofer M; PMU Innovations GmbH, Salzburg, Austria.
  • Hitzl W; Research Office (Biostatistics), Paracelsus Medical University, Salzburg, Austria; Department of Ophthalmology and Optometry, Paracelsus Medical University, Salzburg, Austria.
  • Hinterbuchner L; Department of Internal Medicine, Cardiology and Intensive Care, Paracelsus Medical University, Salzburg, Austria.
  • Seitelberger R; Department of Cardiovascular & Endovascular Surgery, Paracelsus Medical University, Salzburg, Austria.
  • Dinges C; Department of Cardiovascular & Endovascular Surgery, Paracelsus Medical University, Salzburg, Austria. Electronic address: c.dinges@salk.at.
Ann Thorac Surg ; 109(5): 1605-1610, 2020 05.
Article em En | MEDLINE | ID: mdl-31881195
PURPOSE: The aortic root pressurizing device was developed for aortic valve repair surgery. It allows echocardiographic evaluation of the aortic valve during cardioplegic arrest by mimicking diastolic afterload. DESCRIPTION: This polyoxymethylene or polyether ether ketone-based device consists of a sealing cap nut, a sealing ring, a plug screw, and both a filling and a ventilation line. It can be easily connected to any size of aortic Dacron grafts. EVALUATION: The device was tested in 15 porcine hearts using a beating heart biosimulator including hemodynamic and echocardiographic monitoring. Valve competence was analyzed on both the beating and resting heart at 60 and 45 mm Hg root pressure. Aortic insufficiency was surgically induced by leaflet manipulation. Native aortic valves showed either none or trivial aortic insufficiency. After leaflet manipulation echocardiographic proof of valve insufficiency was possible in all specimen. Jet direction was identic in all cases at 60 mm Hg and 93% at 45 mm Hg root pressure. CONCLUSIONS: The aortic root pressurizing device shows highly comparable results of echocardiographic aortic valve evaluation between static and dynamic settings under experimental conditions.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Aorta Torácica / Valva Aórtica / Insuficiência da Valva Aórtica / Próteses Valvulares Cardíacas / Anuloplastia da Valva Cardíaca / Parada Cardíaca Induzida Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Animals Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Aorta Torácica / Valva Aórtica / Insuficiência da Valva Aórtica / Próteses Valvulares Cardíacas / Anuloplastia da Valva Cardíaca / Parada Cardíaca Induzida Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Animals Idioma: En Ano de publicação: 2020 Tipo de documento: Article