Left ventricular inhibition during cardiac resynchronization caused by sensed left-sided ventricular premature complexes.
Barold, S Serge; Kucher, Andreas.
Afiliação
  • Barold SS; Department of Medicine, University of Rochester School of Medicine and Dentistry, Rochester, New York.
  • Kucher A; BIOTRONIK SE & Co. KG, Berlin, Germany.
Pacing Clin Electrophysiol ; 43(2): 240-244, 2020 02.
Article em En | MEDLINE | ID: mdl-31891195
ABSTRACT
A recently published case documented left ventricular (LV) inhibition of a Boston Scientific device by a premature complex (VPC) that was undetected by the right ventricular channel. We have observed a similar response in two patients with a BIOTRONIK cardiac resynchronization device also designed with LV sensing. A spurious response simulating that of the two true cases was also observed in a third patient with a defective LV lead which created isolated false signals. The responses of the BIOTRONIK devices were identical to that of the previously reported findings with the Boston Scientific device. The observations provide insight into the timing function of cardiac resynchronization devices designed with LV sensing.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Complexos Ventriculares Prematuros / Dispositivos de Terapia de Ressincronização Cardíaca Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Complexos Ventriculares Prematuros / Dispositivos de Terapia de Ressincronização Cardíaca Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article