Evaluation of screening effectiveness of hepatitis B surface antigen and anti-HCV rapid test kits in Pakistan.
Waheed, Usman; Abdella, Yetmgeta Eyayou; Saba, Noor E; Arshad, Muhammad; Wazeer, Akhlaaq; Farooq, Ahmad; Usman, Javaid; Arshad, Abida; Zaheer, Hasan Abbas.
Afiliação
  • Waheed U; Safe Blood Transfusion Programme, Ministry of National Health Services, Government of Pakistan, PIMS, Islamabad, Pakistan.
  • Abdella YE; Department of Pathology and Blood Transfusion Services, Shaheed Zulfiqar Ali Bhutto Medical University, PIMS, Islamabad, Pakistan.
  • Saba NE; World Health Organization's Regional Office for Eastern Mediterranean, Cairo, Egypt.
  • Arshad M; Peshawar Regional Blood Centre, Department of Health, Khyber Pakhtunkhwa, Pakistan.
  • Wazeer A; Department of Biological Sciences, International Islamic University, Islamabad, Pakistan.
  • Farooq A; Department of Biological Sciences, International Islamic University, Islamabad, Pakistan.
  • Usman J; Department of Biological Sciences, International Islamic University, Islamabad, Pakistan.
  • Arshad A; Department of Pathology, Army Medical College, Rawalpindi, Pakistan.
  • Zaheer HA; Department of Biology, PMAS Arid Agriculture University, Rawalpindi, Pakistan.
J Lab Physicians ; 11(4): 369-372, 2019.
Article em En | MEDLINE | ID: mdl-31929706
ABSTRACT

OBJECTIVE:

The current study was conducted to evaluate the performance and screening effectiveness of commercially available rapid screening kits in comparison with chemiluminescence immunoassay (CLIA) and polymerase chain reaction (PCR). MATERIALS AND

METHODS:

This single-center, cross-sectional study was conducted at the Department of Pathology and Blood Transfusion Services, Shaheed Zulfiqar Ali Bhutto Medical University, PIMS, Islamabad, from January to April 2019. A total of 10 commercially available immunochromatographic test (ICT) devices and one CLIA kit (LIAISON XL) were tested for their sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and accuracy using 100 positive and 100 negative samples each for HBV and HCV, respectively.

RESULTS:

The sensitivities and specificities of ICT kits for hepatitis B surface antigen were 65% and 70% (Hightop), 67% and 85% (RightSign), 62% and 73% (Wondfo), 70% and 80% (Accu-Chek), 68% and 77% (Fastep), 73% and 85% (Abon), 77% and 83% (ImmuMed), 80% and 90% (Insta-Answer), 67% and 81% (BioCheck), and 72% and 83% CTK Biotech, respectively. Similarly, the sensitivities and specificities of different ICT kits for HCV were 69% and 80% (Hightop), 76% and 83% (RightSign), 69% and 81% (Wondfo), 78% and 79% (Accu-Check), 68% and 68% (Fastep), 63% and 73% (Abon), 71% and 70% (ImmuMed), 79% and 68% (Insta-Answer), 62% and 66% (BioChek), and 69% and 78% CTK Biotech, respectively. The sensitivity and specificity of Diasorin Liaison Murex assay for both HBV and HCV were found to be 100% when compared with PCR. The PPV, NPV and Accuracy were determined accordingly.

CONCLUSION:

Rapid testing ICT devices for both HBV and HCV available in Pakistan were found to have a variable degree of sensitivity and specificity when compared with CLIA and PCR. Comparatively expensive but quality methods are more reliable as compared to rapid devices.