Sensitivity of a Next-Generation NIRS Device to Detect Low Mixed Venous Oxyhemoglobin Saturations in the Single Ventricle Population.
Anesth Analg
; 131(3): e138-e141, 2020 09.
Article
em En
| MEDLINE
| ID: mdl-31985496
ABSTRACT
Regional cerebral oxygenation index (rSO2) based on near-infrared spectroscopy (NIRS) is frequently used to detect low venous oxyhemoglobin saturation (ScvO2). We compared the performance of 2 generations of NIRS devices. Clinically obtained, time-matched cerebral rSO2 and ScvO2 values were compared in infants monitored with the FORE-SIGHT (n = 73) or FORE-SIGHT ELITE (n = 47) by linear regression and Bland-Altman analyses. In both devices, cerebral rSO2 correlated poorly with measured ScvO2 (FORE-SIGHT partial correlation 0.50 [95% confidence interval {CI}, 0.40-0.58]; FORE-SIGHT ELITE partial correlation 0.47 [0.39-0.55]) and mean bias was +8 (standard deviation [SD] 13.2) for FORE-SIGHT and +14 (SD 12.5) for FORE-SIGHT ELITE. When ScvO2 was <30%, rSO2 was <40 in 8% of FORE-SIGHT ELITE readings. Future NIRS should be validated in more hypoxic cohorts.