Patient motion tracking for non-isocentric and non-coplanar treatments via fixed frame-of-reference 3D camera.
Gasparyan, Sergey; Ko, Kyle; Skinner, Lawrie B; Ko, Ryan B; Loo, Billy W; Fahimian, Benjamin P; Yu, Amy S.
Afiliação
  • Gasparyan S; Department of Radiation Oncology, Stanford university, Palo Alto, CA, 94304, USA.
  • Ko K; Department of Radiation Oncology, Stanford university, Palo Alto, CA, 94304, USA.
  • Skinner LB; Department of Radiation Oncology, Stanford university, Palo Alto, CA, 94304, USA.
  • Ko RB; Department of Radiation Oncology, Stanford university, Palo Alto, CA, 94304, USA.
  • Loo BW; Department of Radiation Oncology, Stanford university, Palo Alto, CA, 94304, USA.
  • Fahimian BP; Department of Radiation Oncology, Stanford university, Palo Alto, CA, 94304, USA.
  • Yu AS; Department of Radiation Oncology, Stanford university, Palo Alto, CA, 94304, USA.
J Appl Clin Med Phys ; 21(3): 162-166, 2020 Mar.
Article em En | MEDLINE | ID: mdl-32107845
ABSTRACT

PURPOSE:

As C-arm linac radiation therapy evolves toward faster, more efficient delivery, and more conformal dosimetry, treatments with increasingly complex couch motions are emerging. Monitoring the patient motion independently of the couch motion during non-coplanar, non-isocentric, or dynamic couch treatments is a key bottleneck to their clinical implementation. The goal of this study is to develop a prototype real-time monitoring system for unconventional beam trajectories to ensure a safe and accurate treatment delivery.

METHODS:

An in-house algorithm was developed for tracking using a couch-mounted three-dimensional (3D) depth camera. The accuracy of patient motion detection on the couch was tested on a 3D printed phantom created from the body surface contour exported from the treatment planning system. The technique was evaluated against a commercial optical surface monitoring system with known phantom displacements of 3, 5, and 7 mm in lateral, longitudinal, and vertical directions by placing a head phantom on a dynamic platform on the treatment couch. The stability of the monitoring system was evaluated during dynamic couch trajectories, at speeds between 10.6 and 65 cm/min.

RESULTS:

The proposed monitoring system agreed with the ceiling mounted optical surface monitoring system in longitudinal, lateral, and vertical directions within 0.5 mm. The uncertainty caused by couch vibration increased with couch speed but remained sub-millimeter for speeds up to 32 cm/min. For couch speeds of 10.6, 32.2, and 65 cm/min, the uncertainty ranges were 0.27- 0.73 mm, 0.15-0.87 mm, and 0.28-1.29 mm, respectively.

CONCLUSION:

By mounting a 3D camera in the same frame-of-reference as the patient and eliminating dead spots, this proof of concept demonstrates real-time patient monitoring during couch motion. For treatments with non-coplanar beams, multiple isocenters, or dynamic couch motion, this provides additional safety without additional radiation dose and avoids some of the complexity and limitations of room mounted systems.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Aceleradores de Partículas / Algoritmos / Planejamento da Radioterapia Assistida por Computador / Imagens de Fantasmas / Posicionamento do Paciente / Cabeça / Movimento Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Aceleradores de Partículas / Algoritmos / Planejamento da Radioterapia Assistida por Computador / Imagens de Fantasmas / Posicionamento do Paciente / Cabeça / Movimento Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article