Your browser doesn't support javascript.
loading
The Performance of a Dithiothreitol-Based Diagnostic System in Diagnosing Periprosthetic Joint Infection Compared to Sonication Fluid Cultures and Tissue Biopsies.
Randau, Thomas Martin; Molitor, Ernst; Fröschen, Frank Sebastian; Hörauf, Achim; Kohlhof, Hendrik; Scheidt, Sebastian; Gravius, Sascha; Hischebeth, Gunnar T.
Afiliação
  • Randau TM; Klinik und Poliklinik für Orthopädie und Unfallchirurgie, Universitätsklinikum Bonn.
  • Molitor E; Institut für Medizinische Mikrobiologie, Immunologie und Parasitologie, Universitätsklinikum Bonn.
  • Fröschen FS; Klinik und Poliklinik für Orthopädie und Unfallchirurgie, Universitätsklinikum Bonn.
  • Hörauf A; Institut für Medizinische Mikrobiologie, Immunologie und Parasitologie, Universitätsklinikum Bonn.
  • Kohlhof H; Klinik und Poliklinik für Orthopädie und Unfallchirurgie, Universitätsklinikum Bonn.
  • Scheidt S; Klinik und Poliklinik für Orthopädie und Unfallchirurgie, Universitätsklinikum Bonn.
  • Gravius S; Klinik und Poliklinik für Orthopädie und Unfallchirurgie, Universitätsklinikum Bonn.
  • Hischebeth GT; Orthopädisch-Unfallchirurgisches Zentrum, Universitätsklinikum Mannheim.
Z Orthop Unfall ; 159(4): 447-453, 2021 Aug.
Article em En | MEDLINE | ID: mdl-32392595
ABSTRACT

PURPOSE:

The aim of this study was to evaluate the performance of a commercially available dithiothreitol (DTT) kit for routine use in diagnosing periprosthetic joint infections (PJIs) in comparison to conventional microbiological tissue specimens and sonication procedures in a maximal care hospital.

METHODS:

We applied the DTT system in 40 consecutive cases of revision arthroplasty (23 PJIs and 17 aseptic revisions), with an exchange or a removal of components. The hardware components were split between the DTT system and the conventional sonication procedure. At least three tissue biopsies and a joint fluid specimen were sent for microbiological and histopathological analysis. Data was analysed retrospectively to compare between the different methods.

RESULTS:

Cultures of the DTT fluid showed a sensitivity of 65% and specificity of 100%, as referenced to conventional microbiological cultures. Sonication had better sensitivity (75%) but lower specificity (85%). The categorical agreement of DTT cultures compared to sonication fluid cultures was 78% (31/40). Neither pathogen type, infection duration nor antibiotic pretreatment influenced the accuracy of the DTT, but a low pH in the DTT seemed to be associated with false-negative results.

CONCLUSION:

DTT was inferior in sensitivity when diagnosing PJIs compared to sonication fluid cultures and tissue biopsies. A low pH in the DTT fluid correlated with false-negative results. Nevertheless, the closed system of the DTT kit avoids contamination and false-positive results, and DTT can be an alternative where sonication is not available.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Sonicação / Infecções Relacionadas à Prótese Tipo de estudo: Diagnostic_studies / Observational_studies / Risk_factors_studies Limite: Humans Idioma: En Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Sonicação / Infecções Relacionadas à Prótese Tipo de estudo: Diagnostic_studies / Observational_studies / Risk_factors_studies Limite: Humans Idioma: En Ano de publicação: 2021 Tipo de documento: Article