Your browser doesn't support javascript.
loading
A comparison of the automated blood bank system IH-500 and manual tube method for anti-blood group antibody titration: a quantitative approach.
Yang, John Jeongseok; Seo, Suk Won; Kim, Jin Seok; Chung, Yousun; Kim, Hyungsuk; Ko, Dae-Hyun; Hwang, Sang-Hyun; Heung-Bum, Oh.
Afiliação
  • Yang JJ; Department of Laboratory Medicine, University of Ulsan College of Medicine and Asan Medical Center, Seoul, South Korea.
  • Seo SW; Department of Laboratory Medicine, University of Ulsan College of Medicine and Asan Medical Center, Seoul, South Korea.
  • Kim JS; Department of Laboratory Medicine, University of Ulsan College of Medicine and Asan Medical Center, Seoul, South Korea.
  • Chung Y; Department of Laboratory Medicine, Kangdong Sacred Heart Hospital, Seoul, South Korea.
  • Kim H; Department of Laboratory Medicine, Seoul National University Hospital, Seoul, South Korea.
  • Ko DH; Department of Laboratory Medicine, University of Ulsan College of Medicine and Asan Medical Center, Seoul, South Korea. Electronic address: daehyuni1118@amc.seoul.kr.
  • Hwang SH; Department of Laboratory Medicine, University of Ulsan College of Medicine and Asan Medical Center, Seoul, South Korea.
  • Heung-Bum O; Department of Laboratory Medicine, University of Ulsan College of Medicine and Asan Medical Center, Seoul, South Korea.
Transfus Apher Sci ; 59(4): 102806, 2020 Aug.
Article em En | MEDLINE | ID: mdl-32446633
BACKGROUND: Anti-blood group antibody titers (ABTs) reported in titer values are variable depending on the testing method used. The introduction of new test methods such as automated methods requires proper method comparison. In this study, the automated blood bank system and manual tube method for ABT were compared using a log-transformed dataset to evaluate the alternative statistical approach. METHODS: ABT was conducted using specimens referred for solid organ transplantation. Methods for comparison were conventional manual tube method and IH-500 automated blood bank system using column agglutination (CAT). Criteria for agreement were exact match and 1-titer match. Measured titer values were log-transformed into interval scale for Deming regression analysis. RESULTS: From the comparison of the tube and CAT methods using titer values and the two criteria, the exact and 1-titer match were 15.9-41.5 % and 65.9-97.6 %, respectively. Deming regression was used to demonstrate the presence of both proportional and constant difference between the two methods. CONCLUSION: The method comparison using conventional statistical approaches had limits due to the semi-quantitative value of the test. Log-transformed interval scale values for comparison were useful for interpretation of method comparison datasets. This alternative statistical approach could contribute to a more accurate comparison between assays and standardization of ABT testing.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Sistema ABO de Grupos Sanguíneos / Armazenamento de Sangue Tipo de estudo: Diagnostic_studies / Evaluation_studies Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Sistema ABO de Grupos Sanguíneos / Armazenamento de Sangue Tipo de estudo: Diagnostic_studies / Evaluation_studies Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article