Fully guided implant surgery using Magnetic Resonance Imaging - An in vitro study on accuracy in human mandibles.
Flügge, Tabea; Ludwig, Ute; Winter, Gita; Amrein, Philipp; Kernen, Florian; Nelson, Katja.
Afiliação
  • Flügge T; Department of Oral and Maxillofacial Surgery, Translational Implantology, Medical Center-University of Freiburg, Faculty of Medicine-University of Freiburg, Freiburg, Germany.
  • Ludwig U; Department of Oral and Maxillofacial Surgery, Charité-Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin, Humboldt-Universität zu Berlin, Berlin Institute sof Health, Berlin, Germany.
  • Winter G; Department of Radiology, Medical Physics, Medical Center-University of Freiburg, Faculty of Medicine-University of Freiburg, Freiburg, Germany.
  • Amrein P; Department of Oral and Maxillofacial Surgery, Translational Implantology, Medical Center-University of Freiburg, Faculty of Medicine-University of Freiburg, Freiburg, Germany.
  • Kernen F; Department of Radiology, Medical Physics, Medical Center-University of Freiburg, Faculty of Medicine-University of Freiburg, Freiburg, Germany.
  • Nelson K; Department of Oral and Maxillofacial Surgery, Translational Implantology, Medical Center-University of Freiburg, Faculty of Medicine-University of Freiburg, Freiburg, Germany.
Clin Oral Implants Res ; 31(8): 737-746, 2020 Aug.
Article em En | MEDLINE | ID: mdl-32459868
ABSTRACT

OBJECTIVES:

The objective of this in vitro study was to assess the accuracy of fully guided implant placement following virtual implant planning based on MRI. MATERIAL AND

METHODS:

Sixteen human cadaver hemimandibles with single missing teeth (n = 3), partially edentulous (n = 6) and edentulous situations (n = 7) were imaged using MRI. MRI and optical scans obtained with an intraoral scanner, were imported into an implant planning software. Virtual prosthetic and implant planning were performed regarding hard- and soft-tissue anatomy. Drill guides were manufactured, and fully guided implant placement was performed. Buccal and lingual bone and implant nerve distance were measured by three examiners in preoperative MRI and postoperative CBCT. The implant position was assessed using a software for deviation of implant positions displayed in CBCT and optical scans, respectively.

RESULTS:

MRI displayed relevant structures for implant planning such as cortical and cancellous bone, inferior alveolar nerve and neighboring teeth. Implant planning, CAD/CAM of drill guides and guided implant placement were performed. Deviations between planned and actual implant positions in postoperative CBCT and optical scans were 1.34 mm (SD 0.84 mm) and 1.03 mm (SD 0.46 mm) at implant shoulder; 1.41 mm (SD 0.88 mm) and 1.28 mm (SD 0.52 mm) at implant apex, and 4.84° (SD 3.18°) and 4.21° (SD 2.01°). Measurements in preoperative MRI and postoperative CBCT confirmed the compliance with minimum distances of implants to anatomical structures.

CONCLUSIONS:

Relevant anatomical structures for imaging diagnostics in implant dentistry are displayed with MRI. The accuracy of MRI-based fully guided implant placement in vitro is comparable to the workflow using CBCT.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Implantes Dentários / Cirurgia Assistida por Computador Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Implantes Dentários / Cirurgia Assistida por Computador Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article