Non-enhanced and Non-gated MR Angiography for Robust Visualization of Peripheral Arteries Using Enhanced Acceleration-selective Arterial Spin Labeling (eAccASL).
Magn Reson Med Sci
; 20(3): 312-319, 2021 Sep 01.
Article
em En
| MEDLINE
| ID: mdl-32655087
ABSTRACT
This study aimed to assess the feasibility for applying enhanced acceleration-selective arterial spin labeling (eAccASL) to non-electrocardiogram-gated and non-enhanced peripheral MRA. We compared eAccASL and background suppressed single shot turbo field echo (TFE)-triggered angiography non-contrast-enhanced sequence (BASS TRANCE) required electrocardiographic-gating in eight volunteers and three patients. In the volunteer study, eAccASL demonstrated a comparable arterial visualization compared with BASS TRANCE. In patient observation, the advantages with eAccASL were found in arterial visualization on the collateral vessels and without artifacts affected by arrhythmia events.