Your browser doesn't support javascript.
loading
Variplex™ test system fails to reliably detect SARS-CoV-2 directly from respiratory samples without RNA extraction.
Eckel, Florian; Küsters, Franziska; Drossel, Bernhard; Konert, Markus; Mattes, Hans; Schopf, Stefan.
Afiliação
  • Eckel F; Internal Medicine II, Klinikum rechts der Isar, Technische Universität München, Munich, Germany. florian.eckel@ro-med.de.
  • Küsters F; Medical Department, RoMed Klink Bad Aibling, Harthauser Str. 16, 83043, Bad Aibling, Germany. florian.eckel@ro-med.de.
  • Drossel B; Medical Department, RoMed Klink Bad Aibling, Harthauser Str. 16, 83043, Bad Aibling, Germany.
  • Konert M; Hospital Laboratory, RoMed Klink Bad Aibling, Harthauser Str. 16, 83043, Bad Aibling, Germany.
  • Mattes H; Department of Anesthesia, RoMed Klink Bad Aibling, Harthauser Str. 16, 83043, Bad Aibling, Germany.
  • Schopf S; Hospital Hygiene, RoMed Klinikum Rosenheim, Pettenkoferstr. 10, 83022, Rosenheim, Germany.
Eur J Clin Microbiol Infect Dis ; 39(12): 2373-2377, 2020 Dec.
Article em En | MEDLINE | ID: mdl-32681309
ABSTRACT
Diagnosis of COVID is performed by PCR methods, but their capacity is limited by the requirement of high-level facilities and instruments. The loop-mediated isothermal amplification (LAMP) method has been utilized for the detection of isolated virus-specific RNA. Preliminary data suggest the possibility of isothermal amplification directly from respiratory samples without RNA extraction. All patients admitted to our hospital were screened for SARS-CoV-2 by routine. Respiratory samples were tested by variplex system based on LAMP method directly without RNA extraction and by PCR. Primary endpoint was the false-negative rate of variplex test compared with PCR as gold standard. In 109 patients variplex test and PCR assay were performed simultaneously. Median age was 80 years and male/female ratio was 40/60%. The prevalence of PCR-confirmed COVID diagnosis was 43.1%. Variplex test was positive in 13.8%. False-negative rate of variplex test compared with PCR was 83.0%. The potential of LAMP technology using isolated RNA has been demonstrated impressively by others, and excellent sensitivity and specificity of detecting SARS-CoV-2 has been reported. However, without RNA extraction, the variplex test system failed to reliably detect SARS-CoV-2 directly in respiratory samples.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Pneumonia Viral / RNA Viral / Infecções por Coronavirus / Técnicas de Laboratório Clínico / Técnicas de Amplificação de Ácido Nucleico / Técnicas de Diagnóstico Molecular / Pandemias / Betacoronavirus Tipo de estudo: Diagnostic_studies / Observational_studies / Risk_factors_studies Limite: Aged / Aged80 / Female / Humans / Male País como assunto: Europa Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Pneumonia Viral / RNA Viral / Infecções por Coronavirus / Técnicas de Laboratório Clínico / Técnicas de Amplificação de Ácido Nucleico / Técnicas de Diagnóstico Molecular / Pandemias / Betacoronavirus Tipo de estudo: Diagnostic_studies / Observational_studies / Risk_factors_studies Limite: Aged / Aged80 / Female / Humans / Male País como assunto: Europa Idioma: En Ano de publicação: 2020 Tipo de documento: Article