Analytical evaluation of four faecal immunochemistry tests for haemoglobin.
Piggott, Carolyn; Carroll, Magdalen R R; John, Cerin; O'Driscoll, Shane; Benton, Sally C.
Afiliação
  • Piggott C; NHS Bowel Cancer Screening Programme Southern Hub, Royal Surrey County Hospital, Guildford, England, UK.
  • Carroll MRR; Berkshire and Surrey Pathology Services,Royal Surrey County Hospital, Guildford, England, UK.
  • John C; NHS Bowel Cancer Screening Programme Southern Hub, Royal Surrey County Hospital, Guildford, England, UK.
  • O'Driscoll S; Berkshire and Surrey Pathology Services,Royal Surrey County Hospital, Guildford, England, UK.
  • Benton SC; NHS Bowel Cancer Screening Programme Southern Hub, Royal Surrey County Hospital, Guildford, England, UK.
Clin Chem Lab Med ; 59(1): 173-178, 2020 07 21.
Article em En | MEDLINE | ID: mdl-32692692
ABSTRACT

Objectives:

Faecal immunochemical tests (FIT) for haemoglobin (Hb) are being used in the investigation of colorectal cancer. These tests use antibodies raised to the globin moiety of human Hb. Here, four automated quantitative FIT systems (HM-JACKarc, NS-Prime, OC-Sensor PLEDIA and SENTiFIT 270) are evaluated analytically to confirm whether the performance of the systems meet the manufacturers' claims.

Methods:

Assessment of the analytical performance of the FIT systems was undertaken using Hb lysates, real patient samples and external quality assessment (EQA) samples. This analytical assessment focused on detection characteristics, imprecision, linearity, prozone effect, recovery and carryover.

Results:

All four methods demonstrated good analytical performance, with acceptable within- and between-run imprecision, good recovery of f-Hb and limited carryover of samples. They also all show good linearity across the range of concentrations tested. The results of EQA samples showed different variations from the target values (-52 to 45%), due to the absence of standardisation across the different methods.

Conclusions:

All four systems are fit for purpose and have an analytical performance as documented by their manufacturers.