Your browser doesn't support javascript.
loading
Chronic intracochlear electrical stimulation at high charge densities: reducing platinum dissolution.
Shepherd, Robert K; Carter, Paul M; Enke, Ya Lang; Thompson, Alex; Flynn, Brianna; Trang, Ella P; Dalrymple, Ashley N; Fallon, James B.
Afiliação
  • Shepherd RK; Bionics Institute, St Vincent's Hospital, Melbourne, Australia.
  • Carter PM; Medical Bionics Department, The University of Melbourne, Melbourne, Australia.
  • Enke YL; Cochlear Ltd, Sydney, Australia.
  • Thompson A; Cochlear Ltd, Sydney, Australia.
  • Flynn B; Bionics Institute, St Vincent's Hospital, Melbourne, Australia.
  • Trang EP; Medical Bionics Department, The University of Melbourne, Melbourne, Australia.
  • Dalrymple AN; Bionics Institute, St Vincent's Hospital, Melbourne, Australia.
  • Fallon JB; Bionics Institute, St Vincent's Hospital, Melbourne, Australia.
J Neural Eng ; 17(5): 056009, 2020 10 08.
Article em En | MEDLINE | ID: mdl-32916669
ABSTRACT

OBJECTIVE:

Cochleae of long-term cochlear implant users have shown evidence of particulate platinum (Pt) corroded from the surface of Pt electrodes. The pathophysiological effect of Pt within the cochlea has not been extensively investigated. We previously evaluated the effects of Pt corrosion at high charge densities and reported negligible pathophysiological impact. The present study extends this work by examining techniques that may reduce Pt corrosion.

APPROACH:

Deafened guinea pigs were continuously stimulated for 28 d using biphasic current pulses at extreme charge densities using (i) electrode shorting; (ii) electrode shorting with capacitive coupling (CC); or (iii) electrode shorting with alternating leading phase (AP). On completion of stimulation, cochleae were examined for corrosion product, tissue response, auditory nerve (AN) survival and trace levels of Pt; and electrodes examined for surface corrosion. MAIN

RESULTS:

Pt corrosion was evident at ≥200 µC cm-2 phase-1; the amount dependent on charge density (p< 0.01) and charge recovery technique (p < 0.01); reduced corrosion was apparent using CC. Tissue response increased with charge density (p< 0.007); cochleae stimulated at ≥200 µC cm-2 phase-1 exhibited a vigorous response including a focal region of necrosis and macrophages. Notably, tissue response was not dependent on the charge recovery technique (p = 0.56). Despite stimulation at high charge densities resulting in significant levels of Pt corrosion, there was no stimulus induced loss of ANs.

SIGNIFICANCE:

Significant increases in tissue response and Pt corrosion were observed following stimulation at high charge densities. Charge recovery using CC, and to a lesser extent AP, reduced the amount of Pt corrosion but not the tissue response. Stimulation at change densities an order of magnitude higher than those used when programming cochlear implant recipients in the clinic, produced a vigorous tissue response and corrosion products without evidence of neural loss.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Platina / Potenciais Evocados Auditivos do Tronco Encefálico / Implantes Cocleares / Estimulação Elétrica Limite: Animals Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Platina / Potenciais Evocados Auditivos do Tronco Encefálico / Implantes Cocleares / Estimulação Elétrica Limite: Animals Idioma: En Ano de publicação: 2020 Tipo de documento: Article