Standardized fractal bench test evaluation of coronary stents: Performances in bifurcation lesions treated by the re-proximal optimization technique.
Piriou, Pierre-Guillaume; Bonin, Mickael; Huchet, Francois; Letocart, Vincent; Manigold, Thibaut; Plessis, Julien; Derimay, Francois; Veziers, Joelle; Jordana, Fabienne; Guerin, Patrice.
Afiliação
  • Piriou PG; Department of Cardiology, Nantes University Hospital, Nantes, France.
  • Bonin M; Department of Cardiology, Nantes University Hospital, Nantes, France.
  • Huchet F; Department of Cardiology, Saint-Nazaire Hospital, Saint-Nazaire, France.
  • Letocart V; Department of Cardiology, Nantes University Hospital, Nantes, France.
  • Manigold T; Department of Cardiology, Nantes University Hospital, Nantes, France.
  • Plessis J; Department of Cardiology, Nantes University Hospital, Nantes, France.
  • Derimay F; Department of Interventional Cardiology, Cardiovascular Hospital and INSERM Unit 1060, Lyon, France.
  • Veziers J; INSERM Unit 1229, Regenerative Medicine and Skeleton, Nantes, France.
  • Jordana F; INSERM Unit 1229, Regenerative Medicine and Skeleton, Nantes, France.
  • Guerin P; Department of Cardiology, Nantes University Hospital, Nantes, France.
Catheter Cardiovasc Interv ; 98(1): E9-E17, 2021 07 01.
Article em En | MEDLINE | ID: mdl-32966675
ABSTRACT

BACKGROUND:

Bifurcation lesions in coronary arteries are complex to treat with coronary stents, which are not designed for that purpose and can be unproperly deployed. Moreover, devices are constantly evolving, and so are angioplasty techniques.

OBJECTIVES:

The aim of this study was to determine the performances of different stents in the treatment of bifurcation lesions using the re-proximal optimization technique (rePOT).

METHODS:

Eleven stent platforms were evaluated Xience Sierra (Abbott), Xience Alpine (Abbott), Synergy (Boston), Coroflex Isar (Bbraun), Cobra PzF (Celonova), Ultimaster (Terumo), Resolute Integrity (Medtronic), Resolute Onyx (Medtronic), Optimax (Hexacath), Orsiro (Biotronik), and Absorb (Abbott). Stents were deployed in a silicone fractal bifurcation model using the rePOT. Micro-computed tomography was performed to assess side branch ostium coverage and strut malapposition, as well as the effect of rePOT on stent cell area.

RESULTS:

Our study showed significant differences between stent platforms regarding side branch ostium coverage (p = .002). The Synergy and Cobra PzF stents were the most performant devices to avoid ostium coverage. Strut malapposition varied significantly between devices (p = .008) but the percentage of malapposed struts was relatively low. Significant differences were observed between stents regarding the cell area before (p = .002) and also after rePOT (p = .003), and the increase in cell area caused by rePOT varied considerably between devices (p = .08).

CONCLUSION:

This study highlighted significant differences in the performances of stent platforms deployed in a fractal bifurcation model using rePOT, with a variable impact of the procedure on stent cell area.