Your browser doesn't support javascript.
loading
Dynamic distortion correction for functional MRI using FID navigators.
Wallace, Tess E; Polimeni, Jonathan R; Stockmann, Jason P; Hoge, W Scott; Kober, Tobias; Warfield, Simon K; Afacan, Onur.
Afiliação
  • Wallace TE; Computational Radiology Laboratory, Department of Radiology, Boston Children's Hospital, Boston, Massachusetts, USA.
  • Polimeni JR; Department of Radiology, Harvard Medical School, Boston, Massachusetts, USA.
  • Stockmann JP; Department of Radiology, Harvard Medical School, Boston, Massachusetts, USA.
  • Hoge WS; Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Boston, Massachusetts, USA.
  • Kober T; Department of Radiology, Harvard Medical School, Boston, Massachusetts, USA.
  • Warfield SK; Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Boston, Massachusetts, USA.
  • Afacan O; Department of Radiology, Harvard Medical School, Boston, Massachusetts, USA.
Magn Reson Med ; 85(3): 1294-1307, 2021 03.
Article em En | MEDLINE | ID: mdl-32970869
ABSTRACT

PURPOSE:

To develop a method for slice-wise dynamic distortion correction for EPI using rapid spatiotemporal B0 field measurements from FID navigators (FIDnavs) and to evaluate the efficacy of this new approach relative to an established data-driven technique.

METHODS:

A low-resolution reference image was used to create a forward model of FIDnav signal changes to enable estimation of spatiotemporal B0 inhomogeneity variations up to second order from measured FIDnavs. Five volunteers were scanned at 3 T using a 64-channel coil with FID-navigated EPI. The accuracy of voxel shift measurements and geometric distortion correction was assessed for experimentally induced magnetic field perturbations. The temporal SNR was evaluated in EPI time-series acquired at rest and with a continuous nose-touching action, before and after image realignment.

RESULTS:

Field inhomogeneity coefficients and voxel shift maps measured using FIDnavs were in excellent agreement with multi-echo EPI measurements. The FID-navigated distortion correction accurately corrected image geometry in the presence of induced magnetic field perturbations, outperforming the data-driven approach in regions with large field offsets. In functional MRI scans with nose touching, FIDnav-based correction yielded temporal SNR gains of 30% in gray matter. Following image realignment, which accounted for global image shifts, temporal SNR gains of 3% were achieved.

CONCLUSIONS:

Our proposed application of FIDnavs enables slice-wise dynamic distortion correction with high temporal efficiency. We achieved improved signal stability by leveraging the encoding information from multichannel coils. This approach can be easily adapted to other EPI-based sequences to improve temporal SNR for a variety of clinical and research applications.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Imagem Ecoplanar Limite: Humans Idioma: En Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Imagem Ecoplanar Limite: Humans Idioma: En Ano de publicação: 2021 Tipo de documento: Article