Dosimetric feasibility of brain stereotactic radiosurgery with a 0.35 T MRI-guided linac and comparison vs a C-arm-mounted linac.
Slagowski, Jordan M; Redler, Gage; Malin, Martha J; Cammin, Jochen; Lobb, Eric C; Lee, Brian H; Sethi, Anil; Roeske, John C; Flores-Martinez, Everardo; Stevens, Tynan; Yenice, Kamil M; Green, Olga; Mutic, Sasa; Aydogan, Bulent.
Afiliação
  • Slagowski JM; Radiation and Cellular Oncology, University of Chicago, Chicago, IL, 60637, USA.
  • Redler G; Radiation Oncology, Moffitt Cancer Center, Tampa, FL, 33607, USA.
  • Malin MJ; Radiation Oncology, Langone Medical Center & Laura and Issac Perlmutter Cancer Center, New York University, New York, NY, 10016, USA.
  • Cammin J; Radiation Oncology, Washington University School of Medicine, Barnes-Jewish Hospital, St. Louis, MO, 63110, USA.
  • Lobb EC; Radiation Oncology, St. Elizabeth Hospital, Appleton, WI, 54915, USA.
  • Lee BH; Radiation Oncology, Loyola University Medical Center, Maywood, IL, 60153, USA.
  • Sethi A; Radiation Oncology, Loyola University Medical Center, Maywood, IL, 60153, USA.
  • Roeske JC; Radiation Oncology, Loyola University Medical Center, Maywood, IL, 60153, USA.
  • Flores-Martinez E; Radiation and Cellular Oncology, University of Chicago, Chicago, IL, 60637, USA.
  • Stevens T; Medical Physics, Dalhousie University, Halifax, B3H 4R2, Canada.
  • Yenice KM; Radiation and Cellular Oncology, University of Chicago, Chicago, IL, 60637, USA.
  • Green O; Radiation Oncology, Washington University School of Medicine, Barnes-Jewish Hospital, St. Louis, MO, 63110, USA.
  • Mutic S; Radiation Oncology, Washington University School of Medicine, Barnes-Jewish Hospital, St. Louis, MO, 63110, USA.
  • Aydogan B; Radiation and Cellular Oncology, University of Chicago, Chicago, IL, 60637, USA.
Med Phys ; 47(11): 5455-5466, 2020 Nov.
Article em En | MEDLINE | ID: mdl-32996591
PURPOSE: MRI is the gold-standard imaging modality for brain tumor diagnosis and delineation. The purpose of this work was to investigate the feasibility of performing brain stereotactic radiosurgery (SRS) with a 0.35 T MRI-guided linear accelerator (MRL) equipped with a double-focused multileaf collimator (MLC). Dosimetric comparisons were made vs a conventional C-arm-mounted linac with a high-definition MLC. METHODS: The quality of MRL single-isocenter brain SRS treatment plans was evaluated as a function of target size for a series of spherical targets with diameters from 0.6 cm to 2.5 cm in an anthropomorphic head phantom and six brain metastases (max linear dimension = 0.7-1.9 cm) previously treated at our clinic on a conventional linac. Each target was prescribed 20 Gy to 99% of the target volume. Step-and-shoot IMRT plans were generated for the MRL using 11 static coplanar beams equally spaced over 360° about an isocenter placed at the center of the target. Couch and collimator angles are fixed for the MRL. Two MRL planning strategies (VR1 and VR2) were investigated. VR1 minimized the 12 Gy isodose volume while constraining the maximum point dose to be within ±1 Gy of 25 Gy which corresponded to normalization to an 80% isodose volume. VR2 minimized the 12 Gy isodose volume without the maximum dose constraint. For the conventional linac, the TB1 method followed the same strategy as VR1 while TB2 used five noncoplanar dynamic conformal arcs. Plan quality was evaluated in terms of conformity index (CI), conformity/gradient index (CGI), homogeneity index (HI), and volume of normal brain receiving ≥12 Gy (V12Gy ). Quality assurance measurements were performed with Gafchromic EBT-XD film following an absolute dose calibration protocol. RESULTS: For the phantom study, the CI of MRL plans was not significantly different compared to a conventional linac (P > 0.05). The use of dynamic conformal arcs and noncoplanar beams with a conventional linac spared significantly more normal brain (P = 0.027) and maximized the CGI, as expected. The mean CGI was 95.9 ± 4.5 for TB2 vs 86.6 ± 3.7 (VR1), 88.2 ± 4.8 (VR2), and 88.5 ± 5.9 (TB1). Each method satisfied a normal brain V12Gy  ≤ 10.0 cm3 planning goal for targets with diameter ≤2.25 cm. The mean V12Gy was 3.1 cm3 for TB2 vs 5.5 cm3 , 5.0 cm3 and 4.3 cm3 , for VR1, VR2, and TB1, respectively. For a 2.5-cm diameter target, only TB2 met the V12Gy planning objective. The MRL clinical brain plans were deemed acceptable for patient treatment. The normal brain V12Gy was ≤6.0 cm3 for all clinical targets (maximum target volume = 3.51 cm3 ). CI and CGI ranged from 1.12-1.65 and 81.2-88.3, respectively. Gamma analysis pass rates (3%/1mm criteria) exceeded 97.6% for six clinical targets planned and delivered on the MRL. The mean measured vs computed absolute dose difference was -0.1%. CONCLUSIONS: The MRL system can produce clinically acceptable brain SRS plans for spherical lesions with diameter ≤2.25 cm. Large lesions (>2.25 cm) should be treated with a linac capable of delivering noncoplanar beams.