A novel screening tool to unmask potential interference between S-ICD and left ventricular assist device.
Zormpas, Christos; Eiringhaus, Jörg; Hillmann, Henrike A K; Hohmann, Stephan; Müller-Leisse, Johanna; Schmitto, Jan D; Veltmann, Christian; Duncker, David.
Afiliação
  • Zormpas C; Hannover Heart Rhythm Center, Department of Cardiology and Angiology, Hannover Medical School, Hannover, Germany.
  • Eiringhaus J; Hannover Heart Rhythm Center, Department of Cardiology and Angiology, Hannover Medical School, Hannover, Germany.
  • Hillmann HAK; Hannover Heart Rhythm Center, Department of Cardiology and Angiology, Hannover Medical School, Hannover, Germany.
  • Hohmann S; Hannover Heart Rhythm Center, Department of Cardiology and Angiology, Hannover Medical School, Hannover, Germany.
  • Müller-Leisse J; Hannover Heart Rhythm Center, Department of Cardiology and Angiology, Hannover Medical School, Hannover, Germany.
  • Schmitto JD; Department of Cardiac, Thoracic, Transplant and Vascular Surgery, Hannover Medical School, Hannover, Germany.
  • Veltmann C; Hannover Heart Rhythm Center, Department of Cardiology and Angiology, Hannover Medical School, Hannover, Germany.
  • Duncker D; Hannover Heart Rhythm Center, Department of Cardiology and Angiology, Hannover Medical School, Hannover, Germany.
J Cardiovasc Electrophysiol ; 31(12): 3286-3292, 2020 12.
Article em En | MEDLINE | ID: mdl-33017069
ABSTRACT

INTRODUCTION:

In patients with a left ventricular assist device (LVAD), the subcutaneous implantable cardioverter-defibrillator (S-ICD) can be an alternative to transvenous ICD systems due to reduced risk of systemic infection, which could lead to extraction of the ICD as well as the LVAD. S-ICD eligibility is lower in patients with LVAD than in patients with end-stage heart failure without LVAD. Several reports have shown inappropriate S-ICD therapy in the coexistence of LVAD and S-ICD. The aim of the present study was to evaluate S-ICD eligibility in patients with LVAD using the established electrocardiogram (ECG)-based screening test as well as a novel device-based screening test to identify potentially inappropriate S-ICD sensing in this specific patient cohort. METHODS AND

RESULTS:

The present study included 115 patients implanted with an LVAD. The standard ECG-based screening test and a novel device-based screening test were performed in all patients. Eighty patients (70%) were eligible for S-ICD therapy with the standard ECG-based screening test. Performance of the novel device-based screening test identified device-device interference in 14 of these 80 patients (12%).

CONCLUSION:

Using a novel extended device-based S-ICD screening method, a small number of patients with LVAD deemed eligible for S-ICD with the standard ECG-based screening test exhibit device-device interference. Careful S-ICD screening should be performed in patients with LVAD, who are candidates for S-ICD therapy, to prevent inappropriate sensing or ICD therapy.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Coração Auxiliar / Desfibriladores Implantáveis / Insuficiência Cardíaca Tipo de estudo: Diagnostic_studies / Etiology_studies / Incidence_studies / Observational_studies / Prognostic_studies / Risk_factors_studies / Screening_studies Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Coração Auxiliar / Desfibriladores Implantáveis / Insuficiência Cardíaca Tipo de estudo: Diagnostic_studies / Etiology_studies / Incidence_studies / Observational_studies / Prognostic_studies / Risk_factors_studies / Screening_studies Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article